FDA Adverse Event Malfunction Summary report: N

POMPE A PERFUSION PLUM 360, COMPATIBLE AVEC LE LOGICIEL ICU MEDICAL MEDNET

MDR report key: 10372327 · Received August 6, 2020

Report

Report Number
9615050-2020-00178
Event Type
Malfunction
Date Received
August 6, 2020
Date of Event
July 1, 2020
Report Date
July 20, 2020
Manufacturer
ICU MEDICAL COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K161469
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

H10 - THE DEVICE WAS RECEIVED IN GOOD GENERAL CONDITION. THE DEVICE PASSED A FULL PVT WITHOUT ISSUE. THE CUSTOMER SETTINGS WERE RUN ON 10.50AM (B)(6)2020 . LINE A 550 ML FOR (B)(4) HOURS AT A RATE OF 46 ML/HR. THE DEVICE WAS STOPPED THE FOLLOWING MORNING AT 8.00AM. THE DEVICE COMPLETED THE DELIVERY WITHOUT ISSUE ALARMING N161 LINE A COMPLETED. THE VOLUME DELIVERED WAS APPROXIMATELY 559ML. THE COMPLAINT WAS NOT CONFIRMED THROUGH TESTING. NO PROBABLE CAUSE COULD BE DETERMINED. CLINICAL LOG ANALYSIS INDICATES THAT THE PERCEIVED UNDERDELIVERY WAS DUE TO A LARGE NUMBER OF DELAYS FOR DISTAL OCCLUSION ALARMS AND BACKPRIMING. THERE IS ALSO A LOGGED ALARM FOR A BROKEN DISTAL PRESSURE PIN. THE PIN WAS EXAMINED AND FOUND TO BE IN GOOD CONDITION WITH NO VISIBLE DAMAGE NOTED. THE DEVICE PASSED DISTAL OCCLUSION TESTS WITHOUT ISSUE.

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RETURNED FOR EVALUATION. INVESTIGATION IS PENDING.

Description of Event or Problem · 1

THE EVENT INVOLVED A PLUM 360 PUMP THAT HAD FLOW ISSUES IN THE LAST MONTH. THE DEVICE WAS PROGRAMMED TO INFUSE THE SOLUTION FOR 12 HOURS AND THERE WAS A REPORTED DELAY FOR 3 HOURS, EVEN MORE). THE INFUSION WAS PROGRAMMED ON CHANNEL A: AN UNSPECIFIED SOLUTION OF 550ML FOR 12 HOURS, FROM 20:00H TO 08:00,( FLOW OF 46CC/H). CHANNEL B WAS NOT USED. THE TUBING SET WAS PRIMED AND WAS NOT REPLACED. IT WAS ALSO REPORTED THE CASSETTE DOOR WAS NOT DAMAGED. THERE WAS PATIENT INVOLVEMENT AND DELAY IN THERAPY REPORTED AS THE PUMP HAD TO STAY CONNECTED LONGER AND THE PATIENT DID NOT RECEIVE THE PRODUCT WITHIN THE PRESCRIBED TIME, ALTHOUGH THERE WAS NO NEED FOR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
840121 POMPE A PERFUSION PLUM 360, COMPATIBLE AVEC LE LOGICIEL ICU MEDICAL MEDNET PUMP, INFUSION FRN ICU MEDICAL COSTA RICA LTD.

Patients

Seq Age Sex Outcome Treatment
1 UNSPEC PLUMSET, MFR ICU MEDICAL, INC.| UNSPEC PLUMSET, MFR ICU MEDICAL, INC.