MENTOR SMOOTH ROUND MODERATE PROFILE

MDR report key: 10370327 · Received August 5, 2020

Report

Report Number
1645337-2020-09696
Event Type
Injury
Date Received
August 5, 2020
Date of Event
June 18, 2020
Report Date
July 13, 2020
Manufacturer
MENTOR TEXAS
Product Code
FWM
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION COMPLETED ON AUGUST 25 2020: DURING A VISUAL EVALUATION, AN ANOMALY WAS OBSERVED IN THE ANTERIOR VIEW ON THE DEVICE CONSISTENT WITH A CREASE/FOLD. ADDITIONALLY, A TEAR WAS ALSO OBSERVED WITHIN THE CREASE/FOLD, MEASURING APPROXIMATELY 0.1 CM THE EVALUATION DETERMINED THAT THE LOSS OF SHELL INTEGRITY IS CONSISTENT WITH A CREASE FOLD DEFLATION OF THE IMPLANT SHELL, WHICH IS A KNOWN INHERENT RISK OF SALINE-FILLED MAMMARY PROSTHESIS. AS PART OF OUR QUALITY PROCESS, THE MANUFACTURING RECORDS OF THIS LOT-SERIAL NUMBER WERE REVIEWED AND THE MANUFACTURING STANDARDS WERE MET PRIOR TO THE RELEASE OF THIS LOT. EACH DEVICE IS VISUALLY INSPECTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2020, INDICATED THE DEVICE IDENTITY AS FOLLOW CAT#: 3501680, LOT#: 5910986. A MANUFACTURING RECORD EVALUATION IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION REVIEW COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RUPTURE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) YEAR OLD FEMALE WHO UNDERWENT PRIMARY BREAST AUGMENTATION WITH AN UNKNOWN SALINE PROSTHESIS EXPERIENCED RIGHT SIDED DEFLATION POST PROCEDURE. A PHYSICAL EXAMINATION WAS PERFORMED BY A PHYSICIAN AND DEFLATION WAS DIAGNOSED. AS A RESULT, AN EXPLANT WAS PERFORMED ON (B)(6) 2020.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
834680 MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 5910986

Patients

Seq Age Sex Outcome Treatment
1 42 YR Required Intervention