REFOBACIN BONE CEMENT R 1X40G
Report
- Report Number
- 3006946279-2020-00146
- Event Type
- Malfunction
- Date Received
- August 5, 2020
- Date of Event
- July 20, 2020
- Report Date
- October 12, 2020
- Manufacturer
- BIOMET FRANCE S.A.R.L.
- Product Code
- MBB
- PMA / PMN Number
- K150850
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4, B5, G4, H2, H3, H6, H10. THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER. THEREFORE, IT COULD NOT BE ANALYZED. THE DEVICE MANUFACTURING QUALITY RECORD INDICATES THAT THE RELEASED PRODUCT MET ALL REQUIREMENTS TO PERFORM AS INTENDED. 1 COMPLAINT HAS BEEN RECORDED OVER THE BATCH A746BA1504. ACCORDING TO AVAILABLE DATA, THE MOST PROBABLE ROOT CAUSE IS DUE TO PACKAGING ISSUE (SEALING PROCESS). CORRECTIVE ACTION HAS BEEN INITIATED TO ADDRESS REPORTED ISSUE. A CUSTOMER LETTER WILL BE SEND REGARDING THE REPORTED ISSUE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT HAS BEEN REPORTED THAT THE INNER STERILE PACKAGING WAS FOUND OPENED BEFORE THE OPERATION, NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.
(B)(4). REPORT SOURCE, FOREIGN - EVENT OCCURRED IN (B)(6). THE DEVICE WILL NOT BE RETURNED TO THE MANUFACTURER. THEREFORE IT WILL NOT BE ANALYZED. THE INVESTIGATION IS PENDING. ACCORDING TO AVAILABLE DATA, THE MOST PROBABLE ROOT CAUSE IS DUE TO PACKAGING ISSUE (SEALING PROCESS). CORRECTIVE ACTION HAS BEEN INITIATED TO ADDRESS REPORTED ISSUE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS FOUND THAT THE INNER STERILE PACKAGING WAS OPENED BEFORE THE OPERATION, NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 831953 | REFOBACIN BONE CEMENT R 1X40G | BONE CEMENT, ANTIBIOTIC | MBB | BIOMET FRANCE S.A.R.L. | A746BA1504 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |