OPTILENE 6/0 (0,7) 75CM 2XDR10 CV2 RCP
Report
- Report Number
- 3003639970-2020-00291
- Event Type
- Malfunction
- Date Received
- August 5, 2020
- Date of Event
- July 20, 2020
- Report Date
- September 1, 2020
- Manufacturer
- B. BRAUN SURGICAL, S.A.
- Product Code
- GAW
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED 2,880 UNITS OF THIS CODE-BATCH. THERE ARE 104 UNITS IN OUR STOCK. WE HAVE RECEIVED 26 CLOSED SAMPLES FOR ANALYSIS. WE HAVE TESTED THE NEEDLE ATTACHMENT STRENGTH OF ALL SAMPLES RECEIVED AND THE RESULTS FULFIL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP): 0.25 KGF IN AVERAGE AND 0.135 KGF IN MINIMUM (EP REQUIREMENTS: 0.17 KGF IN AVERAGE AND 0.082 KGF IN MINIMUM) REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL USP/EP AND B.BRAUN SURGICAL REQUIREMENTS. FINAL CONCLUSION: ALTHOUGH THE RESULTS OF THE SAMPLES RECEIVED FULFIL THE SPECIFICATIONS OF EUROPEAN PHARMACOPOEIA/ B. BRAUN SURGICAL SPECIFICATIONS, WE TAKE NOTE OF THIS INCIDENCE IN ORDER TO ASSESS IF NEW OR ADDITIONAL ACTIONS ARE NEEDED. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. CORRECTIVE/PREVENTIVE ACTIONS: ACCORDING TO OUR INTERNAL PROCEDURES, THERE IS NO NEED TO ESTABLISH CORRECTIVE OR PREVENTIVE ACTIONS. NEVERTHELESS, THIS COMPLAINT IS RECORDED FOR TRENDING ANALYSIS TO ASSESS IF ACTIONS ARE NEEDED IN THE FUTURE.
K133890: REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. IF ADDITIONAL INFORMATION BECOMES AVAILABLE A FOLLOW-UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH OPTILENE SUTURE. THE CLIENT REPORTED THAT THE NEEDLE IS DETACHED EASILY FROM THE THREAD. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 832508 | OPTILENE 6/0 (0,7) 75CM 2XDR10 CV2 RCP | CARDIAC SUTURE | GAW | B. BRAUN SURGICAL, S.A. | C3097915 | 119274 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |