FDA Adverse Event Malfunction Summary report: N

OPTILENE 6/0 (0,7) 60CM 2XDRC10 CV2 RCP

MDR report key: 10366222 · Received August 5, 2020

Report

Report Number
3003639970-2020-00293
Event Type
Malfunction
Date Received
August 5, 2020
Date of Event
July 20, 2020
Report Date
September 2, 2020
Manufacturer
B. BRAUN SURGICAL, S.A.
Product Code
GAW
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED 180 UNITS OF THIS CODE-BATCH. THERE ARE 144 UNITS IN OUR STOCK. WE HAVE RECEIVED 33 CLOSED SAMPLES FOR ANALYSIS. WE HAVE TESTED THE NEEDLE ATTACHMENT STRENGTH OF ALL SAMPLES RECEIVED AND THE RESULTS FULFIL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP): 0.29 KGF IN AVERAGE AND 0.148 KGF IN MINIMUM (EP REQUIREMENTS: 0.17 KGF IN AVERAGE AND 0.082 KGF IN MINIMUM). REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL USP/EP AND B.BRAUN SURGICAL REQUIREMENTS. FINAL CONCLUSION: ALTHOUGH THE RESULTS OF THE SAMPLES RECEIVED FULFIL THE SPECIFICATIONS OF EUROPEAN PHARMACOPOEIA/ B. BRAUN SURGICAL SPECIFICATIONS, WE TAKE NOTE OF THIS INCIDENCE IN ORDER TO ASSESS IF NEW OR ADDITIONAL ACTIONS ARE NEEDED. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. CORRECTIVE/PREVENTIVE ACTIONS: ACCORDING TO OUR INTERNAL PROCEDURES, THERE IS NO NEED TO ESTABLISH CORRECTIVE OR PREVENTIVE ACTIONS. NEVERTHELESS, THIS COMPLAINT IS RECORDED FOR TRENDING ANALYSIS TO ASSESS IF ACTIONS ARE NEEDED IN THE FUTURE.

Additional Manufacturer Narrative · 1

K133890: REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. IF ADDITIONAL INFORMATION BECOMES AVAILABLE A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS AN ISSUE WITH OPTILENE SUTURE. THE CLIENT REPORTED THAT THE NEEDLE IS DETACHED EASILY FROM THE THREAD. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
832507 OPTILENE 6/0 (0,7) 60CM 2XDRC10 CV2 RCP CARDIAC SUTURE GAW B. BRAUN SURGICAL, S.A. C3097543 119284

Patients

Seq Age Sex Outcome Treatment
1