AXSYM GENTAMICIN
Report
- Report Number
- 2623532-2008-00011
- Event Type
- Malfunction
- Date Received
- April 30, 2008
- Date of Event
- November 5, 2007
- Report Date
- April 7, 2008
- Manufacturer
- ABBOTT DIAGNOSTICS INTERNATIONAL, LTD.
- Product Code
- LCQ
- PMA / PMN Number
- K935376
- Removal / Correction Number
- 2623532-11/7-07-009-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- UNKNOWN
Narratives
AN INVESTIGATION WAS CONDUCTED FOR THE ABBOTT AXSYM AND TDX/TDXFLX GENTAMICIN REAGENTS. THE ABBOTT AXSYM AND TDX/TDXFLX GENTAMICIN REAGENTS SHARE THE SAME ANTIBODY AND TRACER STOCK CONCENTRATES. A DECREASE IN THE CALIBRATOR A-F SPAN WAS OBSERVED WITH THE TDX/TDXFLX GENTAMICIN REAGENTS RESULTING IN TDX/TDXFLX GENTAMICIN CALIBRATION FAILURES (SPAN LESS THAN MINIMUM SPAN). ALTHOUGH THERE WAS AN APPARENT DECREASE IN THE AXSYM GENTAMICIN REAGENT CALIBRATOR A-F SPAN, THIS ISSUE DID NOT RESULT IN A CALIBRATION FAILURE. NEVERTHELESS, INDIVIDUAL REPLICATES OF THE MEDIUM CONTROL LEVEL OUT OF SPECIFICATION (OOS) FLAGS WERE GENERATED DURING STABILITY FOR TWO TIME POINTS. FOR THE AXSYM GENTAMICIN REAGENT LOTS, SINCE THERE WAS A POTENTIAL THE CALIBRATOR A-F SPAN COULD RESULT IN CALIBRATION FAILURES, A DECISION WAS MADE TO REDUCE THE EXPIRATION DATE FOR THE AXSYM GENTAMICIN REAGENTS, HOWEVER, THE DATA FROM THE INVESTIGATION SUPPORTS THE AXSYM GENTAMICIN REAGENT LOTS WERE NOT IMPACTED AND PERFORMED AS EXPECTED. STABILITY TESTING MONITORED AXSYM GENTAMICIN REAGENT UNTIL THE LAST TEST POINT, WHICH WAS ONE MONTH BEYOND THE REAGENT'S EXPIRATION DATE. THE CALIBRATOR A-F SPAN DID NOT CONTINUE TO DECREASE; INSTEAD, IT INCREASED, SUPPORTING THERE WAS NOT A TRUE DECREASE IN CALIBRATOR A-F SPAN FOR THE AXSYM GENTAMICIN REAGENT. WITH RESPECT TO THE AXSYM GENTAMICIN MEDIUM CONTROL OOS RESULTS, NONE OF THE SUBSEQUENT TEST POINTS SHOWED CONTROL OOS UNTIL THE LAST TEST POINT. STATISTICAL EVAL (ASSAY CAPABILITY) ALONG WITH THESE NEW TEST POINTS SUPPORT THAT REPLICATES OF MEDIUM CONTROL OOS RESULTS WERE RELATED TO THE EXPECTED ASSAY VARIABILITY, THEREFORE, THERE WERE NO PREVENTIVE ACTIONS WARRANTED FOR THE AXSYM GENTAMICIN ASSAY.
THE CUSTOMER CONTACTED ABBOTT AFTER RECEIPT OF ABBOTT'S CUSTOMER LETTER REGARDING AXSYM GENTAMICIN REAGENT PERFORMANCE. THE CUSTOMER STATED THEY HAD BEEN USING THE REAGENT LOT IN QUESTION FOR QUITE A WHILE AND OBSERVED ABBOTT QUALITY CONTROLS QUICKLY DRIFTING OUT OF RANGE HIGH. THE ABBOTT CUSTOMER SERVICE REPRESENTATIVE REQUESTED THE CUSTOMER FAX CALIBRATION AND QUALITY CONTROL DATA TO ABBOTT FOR DATA REVIEW. THE CUSTOMER STATED THE ASSAY PERFORMANCE IS ACCEPTABLE AFTER CHANGING THE LOT OF AXSYM GENTAMICIN REAGENT AND THERE WAS NO IMPACT TO PT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AXSYM GENTAMICIN | FOR THE QUANTITATIVE MEASUREMENT OF GENTAMICIN IN SERUM OR PLASMA | LCQ | ABBOTT DIAGNOSTICS INTERNATIONAL, LTD. | NA | 53190Q100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | AXSYM PLUS ANALYZER |