CELL DYN SAPPHIRE ANALYZER
Report
- Report Number
- 2919069-2008-00521
- Event Type
- Malfunction
- Date Received
- April 29, 2008
- Date of Event
- February 13, 2008
- Report Date
- April 10, 2008
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISION/CELL DYN
- Product Code
- GKZ
- PMA / PMN Number
- K051215
- Removal / Correction Number
- 2919069-4/15/08-0003-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- UNKNOWN
Narratives
THE INVESTIGATION FOUND AN ISSUE WITH THE CELL-DYN SAPPHIRE LABORATORY INFO SYSTEM (LIS) / MIDDLEWARE INTERFACE CHANGE SPECIFICATION, LIST #08H09-01. THESE CHANGES ARE OPTIONAL, THIS MEANS THAT, IF CUSTOMERS DO NOT WANT THE ADD'L FEATURES ACTIVATED, THEIR LIS WILL NOT CHANGE. THE LIS TRANSMISSION FOR VERSION R2-5 SOFTWARE CONTAINS A CHANGE IN THE ORDER OF RESULTS TRANSMITTED TO THE LIS, WHICH HAS CREATED THE ISSUE WITH SOME EXISTING LIS IF THE CUSTOMER DOES NOT CORRECTLY MAP THE LIS BY THE NUMERICAL RESULT LABEL PRIOR TO UPGRADING TO R2-5 SOFTWARE. BASED ON THE ABOVE INVESTIGATION, A PRODUCT DEFICIENCY WAS IDENTIFIED FOR THE CELL-DYN SAPPHIRE LIS / MIDDLEWARE INTERFACE SPECIFICATION, LIST # 08H09-01. A REVIEW OF THE DOMESTIC AND INTERNATIONAL COMPLAINTS FOR THE MONTHS OF 2007 THROUGH 2008 DID NOT FIND AN ADVERSE TREND ON THE LIST BASE HEM SOFTWARE FOR THIS ISSUE. THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE CUSTOMER CLAIMED THAT THE ORDER OF THE RESULTS TRANSMISSION OF THE CELL-DYN SAPPHIRE LAB INFO SYSTEM (LIS) SOFTWARE HAS CHANGED WITHOUT NOTICE WHEN SWITCHING FROM REVISION 2.4 TO REVISION 2.5. THIS CHANGE WAS NOT MENTIONED IN THE NEW REVISION'S SPECIFICATION. THERE WAS NO IMPACT TO PT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CELL DYN SAPPHIRE ANALYZER | AUTOMATED HEMATOLOGY ANALYZER | GKZ | ABBOTT DIAGNOSTICS DIVISION/CELL DYN | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | CELL-DYN SAPPHIRE UPGRADED KIT. LIST # 9H34-04 |