FDA Adverse Event Malfunction Summary report: N

IMP,TSV,4.1MM,SBM,8

MDR report key: 10360749 · Received August 4, 2020

Report

Report Number
0002023141-2020-01170
Event Type
Malfunction
Date Received
August 4, 2020
Date of Event
June 23, 2020
Report Date
October 29, 2020
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019263
PMA / PMN Number
K072589
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). ONE TAPERED SCREW-VENT IMPLANT (TSV4B8) WAS RETURNED FOR INVESTIGATION. VISUAL EVALUATION OF THE AS RETURNED PRODUCT IDENTIFIED SIGNS OF WEAR AROUND THE INTERNAL HEX DUE TO USAGE. FUNCTIONAL TESTING WAS PERFORMED USING AN APPLICABLE IN-HOUSE ABUTMENT. THE DEVICE FAILED TO FULLY ASSEMBLE WITH THE ABUTMENT. PRE-EXISTING CONDITION NOTED ON THE PER WAS HIGH BONE DENSITY ¿ TYPE I. THE DEVICE WAS BEING PLACED ON TOOTH # 36 (FDI) WHEN THE INCIDENT OCCURRED. DEVICE HISTORY RECORD (DHR) REVIEW FOR THE LOT (1221590) HAD REVEALED NO DEVIATIONS NOR NON-CONFORMANCES WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. ALL PRODUCTS WERE CONFORMING AT THE TIME THEY LEFT ZIMMER BIOMET. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE LOT (1221590) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT ABOUT NONCONFORMING PRODUCTS WERE IDENTIFIED. THEREFORE, BASED ON THE AVAILABLE INFORMATION AND FUNCTIONAL TESTING, DEVICE MALFUNCTION DID OCCUR AND THE REPORTED EVENT WAS CONFIRMED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. FIRST/GIVEN NAME: UNKNOWN / NOT PROVIDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANT HEX CONNECTION WAS DAMAGED DURING IMPLANT PLACEMENT, DOCTOR REMOVED THE IMPLANT AND PLACED ANOTHER ONE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
829343 IMP,TSV,4.1MM,SBM,8 DENTAL IMPLANT DZE ZIMMER DENTAL TSV4B8 1221590 00889024019263

Patients

Seq Age Sex Outcome Treatment
1 58 YR