FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.3ML 30GA 8MM 7BAG 420CAS JP

MDR report key: 10359529 · Received August 3, 2020

Report

Report Number
1920898-2020-00978
Event Type
Malfunction
Date Received
August 3, 2020
Date of Event
July 13, 2020
Report Date
August 11, 2020
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: CUSTOMER RETURNED (21) 3/10CC, 8MM, 30G SYRINGES IN SEALED POLY BAGS FROM LOT # 9210505. CUSTOMER STATES THAT WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE HUB. ALL RETURNED SYRINGES WERE TESTED AND NO HUB-NEEDLE ASSEMBLY SEPARATES WHEN THE SHIELD WAS REMOVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 9210505 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT DID NOT PERTAIN TO MY COMPLAINT. ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT 21 SYRINGE 0.3ML 30GA 8MM 7BAG 420CAS JP EXPERIENCED HUB SEPARATION FROM THE DEVICE PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE HUB.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT 21 SYRINGE 0.3ML 30GA 8MM 7BAG 420CAS JP EXPERIENCED HUB SEPARATION FROM THE DEVICE PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE HUB.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
818857 SYRINGE 0.3ML 30GA 8MM 7BAG 420CAS JP SYRINGE FMF BD MEDICAL - DIABETES CARE 9210505

Patients

Seq Age Sex Outcome Treatment
1 Other