FDA Adverse Event Malfunction Summary report: N

RUSCH TRULED HANDLE, MEDIUM, EUROPE

MDR report key: 10356246 · Received August 3, 2020

Report

Report Number
8030121-2020-00135
Event Type
Malfunction
Date Received
August 3, 2020
Date of Event
July 15, 2020
Report Date
July 29, 2020
Manufacturer
TELEFLEX MEDICAL
Product Code
CCW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THE SAMPLE WAS RETURNED FOR EVALUATION AND SENT TO THE MANUFACTURING SITE FOR INVESTIGATION. THE MANUFACTURING SITE REPORTS THAT THE CARTRIDGE WAS "RESPONDING RED AND GREEN LIGHT WHEN IT HAD CONNECTED WITH THE POWER SOURCE LIKE CHARGER WITH ELECTRICITY". THE MANUFACTURING SITE ALSO REPORTS THAT IT WAS OBSERVED THAT THERE WAS A THICK FORMATION INSIDE OF THE INNER CARTRIDGE INDICATING AN INCORRECT STERILIZATION OR CLEANING PROCESS. THE DEVICE HISTORY RECORD OF LOT OF THE SAMPLE RECEIVED (190601) WAS REVIEWED AND NO ISSUE THAT COULD HAVE CONTRIBUTED TO THE REPORTED FAILURE WAS NOTED. THE DEVICE WAS MANUFACTURED ACCORDING TO RELEASE SPECIFICATION. THE MANUFACTURING SITE REPORTS THE "FAILURE CANNOT BE JUSTIFIED AS DURING INVESTIGATION WE NOTICED TRACES OF CHEMICAL /DISINFECTANT ON THE SURFACE ON THE INNER CARTRIDGES AND DEFINITELY DISINFECTANT /CHEMICAL HAD ENTERED ON THE INNER SURFACE AND THERE IS POSSIBILITY THAT DUE TO CHEMICAL EXPOSURE INTERNAL ELECTRICAL CIRCUIT OF LED HAS GOT FUSED. AS PER THE IFU INSTRUCTIONS THE OUTER HANDLE MAY BE AUTOCLAVED OR STERILIZED VIA STERIS OR STERRAD AND THE INNER CARTRIDGE IS NOT SUITABLE FOR STERILIZATION BUT DURING INVESTIGATION WE HAVE TRACES THE CHEMICAL AGENT LAYER ON INNER CARTRIDGE.

Description of Event or Problem · 0

IT WAS REPORTED THE LED LIGHT DID NOT WORK PRIOR TO PATIENT USE.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THE LED LIGHT DID NOT WORK PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
822643 RUSCH TRULED HANDLE, MEDIUM, EUROPE LARYNGOSCOPE, RIGID CCW TELEFLEX MEDICAL 190301

Patients

Seq Age Sex Outcome Treatment
1