RUSCH TRULED HANDLE, MEDIUM, EUROPE
Report
- Report Number
- 8030121-2020-00135
- Event Type
- Malfunction
- Date Received
- August 3, 2020
- Date of Event
- July 15, 2020
- Report Date
- July 29, 2020
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- CCW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SAMPLE WAS RETURNED FOR EVALUATION AND SENT TO THE MANUFACTURING SITE FOR INVESTIGATION. THE MANUFACTURING SITE REPORTS THAT THE CARTRIDGE WAS "RESPONDING RED AND GREEN LIGHT WHEN IT HAD CONNECTED WITH THE POWER SOURCE LIKE CHARGER WITH ELECTRICITY". THE MANUFACTURING SITE ALSO REPORTS THAT IT WAS OBSERVED THAT THERE WAS A THICK FORMATION INSIDE OF THE INNER CARTRIDGE INDICATING AN INCORRECT STERILIZATION OR CLEANING PROCESS. THE DEVICE HISTORY RECORD OF LOT OF THE SAMPLE RECEIVED (190601) WAS REVIEWED AND NO ISSUE THAT COULD HAVE CONTRIBUTED TO THE REPORTED FAILURE WAS NOTED. THE DEVICE WAS MANUFACTURED ACCORDING TO RELEASE SPECIFICATION. THE MANUFACTURING SITE REPORTS THE "FAILURE CANNOT BE JUSTIFIED AS DURING INVESTIGATION WE NOTICED TRACES OF CHEMICAL /DISINFECTANT ON THE SURFACE ON THE INNER CARTRIDGES AND DEFINITELY DISINFECTANT /CHEMICAL HAD ENTERED ON THE INNER SURFACE AND THERE IS POSSIBILITY THAT DUE TO CHEMICAL EXPOSURE INTERNAL ELECTRICAL CIRCUIT OF LED HAS GOT FUSED. AS PER THE IFU INSTRUCTIONS THE OUTER HANDLE MAY BE AUTOCLAVED OR STERILIZED VIA STERIS OR STERRAD AND THE INNER CARTRIDGE IS NOT SUITABLE FOR STERILIZATION BUT DURING INVESTIGATION WE HAVE TRACES THE CHEMICAL AGENT LAYER ON INNER CARTRIDGE.
IT WAS REPORTED THE LED LIGHT DID NOT WORK PRIOR TO PATIENT USE.
(B)(4).
IT WAS REPORTED THE LED LIGHT DID NOT WORK PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 822643 | RUSCH TRULED HANDLE, MEDIUM, EUROPE | LARYNGOSCOPE, RIGID | CCW | TELEFLEX MEDICAL | 190301 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |