FDA Adverse Event Injury Summary report: N

REVERSE SHOULDER SYSTEM GLENOID POLYAXIAL LOCKING SCREW - L22

MDR report key: 10349979 · Received July 31, 2020

Report

Report Number
3005180920-2020-00485
Event Type
Injury
Date Received
July 31, 2020
Date of Event
July 9, 2020
Report Date
July 31, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
PHX
UDI-DI
07630040706469
PMA / PMN Number
K170452
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 21 JULY 2020: LOT 174282: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 13-NOV-2018. EXPIRATION DATE: 2022-11-02. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. 2 DEVICES OF THIS LOT INVOLVED IN THE COMPLAINT. OTHER DEVICES INVOLVED ON THE EVENT: REVERSE SHOULDER SYSTEM 04.01.0157 GLENOID POLYAXIAL LOCKING SCREW - L14 LOT. 174280 (K170452). BATCH REVIEW PERFORMED ON 21 JULY 2020: LOT 174280: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 13-NOV-2017. EXPIRATION DATE: 2022-11-01. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. REVERSE SHOULDER SYSTEM 04.01.0161 GLENOID POLYAXIAL LOCKING SCREW - L30 LOT. 174284 (K170452). BATCH REVIEW PERFORMED ON 21 JULY 2020: LOT 174284: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12-OCT-2017. EXPIRATION DATE: 2022-10-08. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. REVERSE SHOULDER SYSTEM 04.01.0172 GLENOSPHERE 36X27 LOT. 163802 (K170452). BATCH REVIEW PERFORMED ON 21 JULY 2020: LOT 163802: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 20-DEC-2016. EXPIRATION DATE: 2011-11-06. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. REVERSE SHOULDER SYSTEM 04.01.0110 HUMERAL REVERSE METAPHYSIS +0MM/0° LOT. 174648 (K170452). BATCH REVIEW PERFORMED ON 21 JULY 2020: LOT 174648: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 06-DEC-2017. EXPIRATION DATE: 2022-11-27. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. REVERSE SHOULDER SYSTEM 04.01.0119 HUMERAL REVERSE HC LINER 36/+0MM LOT. 173420 (K170452) BATCH REVIEW PERFORMED ON 21 JULY 2020: LOT 173420: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 04-JUL-2017. EXPIRATION DATE: 2022-06-18. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. REVERSE SHOULDER SYSTEM 04.01.0154 GLENOID BASEPLATE Ø27X15 LOT. 163781 (K170452). BATCH REVIEW PERFORMED ON 21 JULY 2020: LOT 163781: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 15-DEC-2016. EXPIRATION DATE: 2021-08-21. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

DURING THE FIRST SURGERY (2018) THERE WAS AN INTRA-OP CAUDAL GLENOID FRACTURE DURING THE IMPLANTATION OF AN INVERSE SHOULDER IMPLANT. THE SURGEON REPORTED TO AGENT THAT HE HAS IN SUCH CASES 2 OPTIONS: 1) GO FOR A HEMI PROTHESIS OR 2) GO FOR AN INVERSE PROTHESIS BUT IMPLANT IT TOO CRANIAL AND THEN LATER REVISE THE GLENOID SIDE FOR A ANATOMIC POSITION AFTER FRACTURE HEALING AND IN CASE OF PROBLEM. THE SURGEON HAS CHOSEN OPTION 2 TO GO ON. THIS REVISION OF TODAY WAS THE REVISION OF THE OPTION 2 - A "PLANNED REVISION". THE SURGEON TOLD ME THAT THE REVISION IS NOT IMPLANT RELATED. DIAPHYSIS WAS NOT REMOVED. ALL OTHER COMPONENTS WERE REMOVED AND SENT TO SONIFICATION (WHY NOT AVAILABLE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
813661 REVERSE SHOULDER SYSTEM GLENOID POLYAXIAL LOCKING SCREW - L22 GLENOID POLYAXIAL LOCKING SCREW PHX MEDACTA INTERNATIONAL SA 04.01.0159 174282 07630040706469

Patients

Seq Age Sex Outcome Treatment
1 79 YR Required Intervention