E1704 DENTAL CHAIR
Report
- Report Number
- 1017522-2020-00011
- Event Type
- Malfunction
- Date Received
- July 31, 2020
- Date of Event
- July 16, 2020
- Report Date
- July 31, 2020
- Manufacturer
- KAVO DENTAL TECHNOLOGIES, LLC
- Product Code
- NRU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
AN AUTHORIZED DEALER SERVICE TECHNICIAN PROVIDED A PHOTO OF THE E1704 DENTAL CHAIR TO KAVO DENTAL TECHNOLOGIES. UPON REVIEW OF THE PROVIDED PHOTO, IT APPEARS THE LOWER TRUSS AREA OF THE DENTAL CHAIR HAS BROKEN. THE EXACT NATURE AND CAUSE OF THE DAMAGE TO THE LOWER TRUSS AREA COULD NOT BE DETERMINED FROM THE PHOTOGRAPH. THERE WAS NO REPORT OF INJURY OR IMPACT TO PATIENT CARE, AND THE AUTHORIZED DEALER SERVICE TECHNICIAN WAS DIRECTED BY THE KAVO DENTAL TECHNOLOGIES, LLC, TECHNICAL SERVICE DEPARTMENT TO ADVISE THE CUSTOMER TO DISCONTINUE USE OF THE DENTAL CHAIR. A FOLLOW UP CALL TO THE DOCTOR'S OFFICE ON 29-JULY-2020 CONFIRMED THROUGH THE OFFICE MANAGER THAT THE CHAIR HAS BEEN REMOVED FROM PATIENT SERVICE. THE DENTAL CHAIR WAS MANUFACTURED IN JUNE 2011 AND IS OVER 8 YEARS OLD AND IS PAST THE EXPECTED LIFE OF THE DEVICE. THIS CONCLUDES THE INVESTIGATION AND THIS REPORT.
AN ALLEGED MALFUNCTION OF MODEL E1704 DENTAL CHAIR SERIAL NUMBER (B)(4) WAS REPORTED TO KAVO DENTAL TECHNOLOGIES, LLC ON (B)(6) 2020 BY AN AUTHORIZED DEALER SERVICE TECHNICIAN. THE REPORT ALLEGED THAT THE MODEL E1704 DENTAL CHAIR TRUSS HAD BUCKLED AND ALLOWING THE CHAIR TO LEAN FORWARD. THERE WAS NO REPORT OF INJURY OR IMPACT TO CUSTOMER CARE AND THE AUTHORIZED DEALER SERVICE TECHNICIAN WAS DIRECTED BY THE KAVO DENTAL TECHNOLOGIES, LLC, TECHNICAL SERVICE DEPARTMENT TO ADVISE THE CUSTOMER TO DISCONTINUE USE OF THE CHAIR. THE DENTAL CHAIR WAS MANUFACTURED IN JUNE 2011 AND IS OVER 8 YEARS OLD AND IS PAST THE EXPECTED LIFE OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 813893 | E1704 DENTAL CHAIR | DENTAL CHAIR | NRU | KAVO DENTAL TECHNOLOGIES, LLC | E1704 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |