FDA Adverse Event Malfunction Summary report: N

PICCOLO COMPREHENSIVE METABOLIC PANEL

MDR report key: 10347817 · Received July 30, 2020

Report

Report Number
2939693-2020-00082
Event Type
Malfunction
Date Received
July 30, 2020
Date of Event
September 27, 2019
Report Date
July 24, 2020
Manufacturer
ABAXIS, INC.
Product Code
CHJ
PMA / PMN Number
K934592
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(6) 2019, ABAXIS RECEIVED A CALL FROM THE CUSTOMER LAB STATING THAT THE DEVICE YIELDED MULTIPLE UNEXPECTED HIGH CHLORIDE RESULTS ON MULTIPLE PATIENTS. THE RECOMMENDED REFERENCE RANGE FOR THE PICCOLO SYSTEM IS 98-108MMOL/L. THE LAB RAN CONTROLS ON THE LOT ON (B)(6) 2019 AND BOTH LEVELS WERE IN RANGE. A REVIEW OF THE BATCH RECORD SHOWED THAT THE LOT MET ALL RELEASE CRITERIA. THE LAB RETURNED THEIR ANALYZER FOR REPAIR. IT WAS OBSERVED THAT OCCASIONAL TEMPERATURE SPIKES ABOVE 38 DEGREES CELSIUS WERE OCCURRING. THE HEATER PLATES WERE REPLACE TO RESOLVE THE REPORTED PROBLEM. THE TEMPERATURE WAS RECALIBRATED AND THE DEVICE WAS THOROUGHLY CLEANED. THE DEVICE WAS SENT BACK TO THE CUSTOMER SITE WHERE IT REMAINS. NO FURTHER ACTIONS WERE REQUIRED. THIS CALL WAS ORIGINALLY ONE ISSUE THAT WAS SPLIT DUE TO MULTIPLE PANEL LOTS (9215AB7, 9272AB1 9242AB1). THIS MDR ADDRESSES 9242AB1. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER LAB REPORTED THAT THE DEVICE YIELDED MULTIPLE UNEXPECTED HIGH CHLORIDE RESULTS ON MULTIPLE PATIENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
808198 PICCOLO COMPREHENSIVE METABOLIC PANEL PICCOLO METABOLIC PANEL CHJ ABAXIS, INC. 9242AB1

Patients

Seq Age Sex Outcome Treatment
1