FDA Adverse Event Injury Summary report: Y

EDWARDS ESHEATH INTRODUCER SET

MDR report key: 10344330 · Received July 30, 2020

Report

Report Number
2015691-2020-12819
Event Type
Injury
Date Received
July 30, 2020
Date of Event
May 29, 2020
Report Date
July 27, 2020
Manufacturer
EDWARDS LIFESCIENCES
Product Code
NPT
PMA / PMN Number
P140031
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT TO ADD NOE STATEMENT IN B5 SECTION, D5, F10, H6.

Description of Event or Problem · 0

THIS REPORT SUMMARIZE <NOE> 30 </NOE> EVENTS OF BLEEDING AT ACCESS SITE SERIOUS INJURY EVENTS FOR THE SAPIEN 3 FOR (B)(6) OF 2020.

Additional Manufacturer Narrative · 1

EXEMPTION NUMBER E2016006, THIS REPORT SUMMARIZES 30 EVENTS OF BLEEDING AT ACCESS SITE FOR THE SAPIEN 3 VALVE IN THE AORTIC POSITION. THE ¿TIME TO EVENT¿ (TTE, IN DAYS) FOR THIS EVENT WAS 0.06. THE DEVICE IDENTIFICATION (DI) NUMBERS FOR EDWARDS ESHEATH INTRODUCER SET ARE (B)(4). THE INSTRUCTIONS FOR USE (IFU) LIST HEMORRHAGE REQUIRING TRANSFUSION OR INTERVENTION AS A POTENTIAL RISK ASSOCIATED WITH THE OVERALL TAVR PROCEDURE, INCLUDING POTENTIAL ACCESS COMPLICATIONS ASSOCIATED WITH STANDARD CARDIAC CATHETERIZATION. THE MAJORITY OF TRANSFEMORAL ACCESS RELATED BLEEDING COMPLICATIONS ARE RELATED TO DISEASED ILEO-FEMORAL VESSELS AND/OR PROCEDURAL TECHNIQUE DURING THE INSERTION OR REMOVAL OF THE SHEATH AND/OR DILATORS. THERE ARE CASES WHERE THE BLEEDING IS SIGNIFICANT AND MORE COMPLEX INTERVENTION OR TRANSFUSION IS REQUIRED TO TREAT/PREVENT A PERMANENT INJURY FROM OCCURRING. THE THV PHYSICIAN TRAINING MANUALS INSTRUCT ON PROCEDURAL CONSIDERATIONS FOR SHEATH INSERTION WITH REGARDS TO PROPER SCREENING CRITICAL TO REDUCING VASCULAR COMPLICATIONS. THE IFU CONTRAINDICATES PATIENTS WITH ILIO-FEMORAL VESSEL CHARACTERISTICS THAT WOULD PRECLUDE SAFE PLACEMENT OF SHEATHS SUCH AS SEVERE OBSTRUCTIVE CALCIFICATION OR SEVERE TORTUOSITY. PRE-PROCEDURE SCREENING AND ASSESSMENT OF THE FEMORAL/ILIAC ARTERY INTERNAL DIAMETERS WILL ENABLE THE CLINICIAN TO DETERMINE IF THE VALVE CAN BE DELIVERED TRANSFEMORALLY. IN THIS CASE, SPECIFIC PROCEDURAL DETAILS ARE NOT AVAILABLE TO DETERMINE POTENTIAL CONTRIBUTING FACTORS TO THE EVENT, OR IF THE EVENT IS RELATED TO AN EDWARDS DEVICE; HOWEVER, A TRANSFUSION OF 2 UNITS OR MORE OF BLOOD WAS REPORTED IN THE REGISTRY. BY VARC DEFINITION, THE MEET CRITERIA FOR MAJOR VASCULAR COMPLICATION, THEREFORE THE EVENT IS BEING REPORTED. THE IFU AND TRAINING MANUALS HAVE BEEN REVIEWED AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED AT THIS TIME.

Description of Event or Problem · 1

THV/TVT REGISTRY SUMMARY REPORTING FOR ADVERSE EVENT SUBMISSION FOR (B)(6) 2020 DATA EXTRACT FOR AORTIC SERIOUS INJURIES FOR THE SAPIEN 3 VALVE. THIS REPORT SUMMARIZES 30 EVENTS OF BLEEDING AT ACCESS SITE IN THE AORTIC POSITION SERIOUS INJURY EVENTS FOR THE SAPIEN 3 TRANSCATHETER HEART VALVE FOR (B)(6) 2020. THE AGE RANGE FOR THESE EVENTS IS FROM 58-93. THE BREAKDOWN FOR GENDER IS AS FOLLOWS: 18 FEMALES AND 12 MALES. (B)(6) 2020 DATA EXTRACT INCLUDES DATA PROVIDED BY ACC FOR Q4 2019 (OCTOBER 1 ¿ DECEMBER 31).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
811096 EDWARDS ESHEATH INTRODUCER SET AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED NPT EDWARDS LIFESCIENCES ESHEATH UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 77 YR Other