FDA Adverse Event Injury Summary report: N

PUMP MMT-712LNAB PRDGM INS BL EN US LN

MDR report key: 1033884 · Received April 24, 2008

Report

Report Number
2032227-2008-00709
Event Type
Injury
Date Received
April 24, 2008
Date of Event
April 11, 2008
Report Date
April 13, 2008
Manufacturer
MEDTRONIC MINIMED
Product Code
LZG
PMA / PMN Number
K031390
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT SHE WAS HOSPITALIZED FOR DIABETIC KETOACIDOSIS. NO BLOOD GLUCOSE READING WAS REPORTED. TROUBLESHOOTING WAS PERFORMED AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP ALSO PASSED THE PRIME AND HIGH PRESSURE TESTS. ALSO FOUND SEVERAL NO DELIVERY ALARMS IN THE ALARM HISTORY, ALL FROM THE DAY OF THE HOSPITALIZATION. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-712LNAB PRDGM INS BL EN US LN INFUSION PUMP LZG MEDTRONIC MINIMED MMT-712LNAB

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization