FDA Adverse Event Injury Summary report: N

ZENITH AAA ILIAC LEG GRAFT

MDR report key: 1033873 · Received April 25, 2008

Report

Report Number
1820334-2008-00240
Event Type
Injury
Date Received
April 25, 2008
Date of Event
March 27, 2008
Report Date
March 27, 2008
Manufacturer
COOK INCORPORATED
Product Code
MIH
PMA / PMN Number
P020018
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: ZENITH DEVICES ARE EACH SHIPPED WITH AN INSTRUCTIONS FOR USE (IFU) BOOKLET THAT LISTS THE ANATOMICAL REQUIREMENTS, WARNINGS, PRECAUTIONS, AND THE CORRECT DEPLOYMENT PROCEDURE. THE IFU ALSO ADVISES ON APPROPRIATE FOLLOW-UP SO THAT LEAKS, SHOULD THEY OCCUR, CAN BE IDENTIFIED AND ADDRESSED BEFORE CAUSING CLINICAL PROBLEMS. AN INTERNAL CLINICAL REVIEW INDICATED THAT THE EVENT WAS CAUSED BY ANATOMICAL CHANGES OVER TIME. THE DEVICE REMAINS IMPLANTED AND NO FILMS WERE PROVIDED TO ASSIST IN THIS INVESTIGATION.

Description of Event or Problem · 1

A MALE UNDERWENT INITIAL AAA REPAIR IN 2004. ONE MAIN BODY GRAFT, TWO ILIAC LEG GRAFTS, AND ONE LEG EXTENSION GRAFT WERE PLACED. THE PHYSICIAN WAS AWARE DURING THE FIRST PROCEDURE THAT THE LEFT ILIAC LEG GRAFT MAY NOT SEAL DUE TO THE LARGE DIAMETER OF THE PATIENT'S ILIAC ARTERY. AT THAT TIME, THE PHYSICIAN DECIDED NOT TO EXTEND THE GRAFT INTO THE EXTERNAL ILIAC, THUS COVERING THE INTERNAL ILIAC. INSTEAD, HE DECIDED TO MONITOR THE SIZE OF THE ILIAC AND THE AAA WITH FOLLOW-UP CT SCANS OVER SEVERAL MONTHS. THE CT SCANS REVEALED THE ILIAC AND AAA WERE CONTINUING TO GROW. THERE WAS A TYPE I ENDOLEAK. THEREFORE, THE PHYSICIAN DECIDED TO EMBOLIZE THE INTERNAL ILIAC AND THEN EXTEND THE GRAFT INTO THE EXTERNAL ILIAC BY PLACING ANOTHER ILIAC LEG GRAFT TO PREVENT RETROGRADE FLOW INTO THE ILIAC AND AAA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ZENITH AAA ILIAC LEG GRAFT MIH ENDOVASCULAR GRAFT MIH COOK INCORPORATED NA 1403507

Patients

Seq Age Sex Outcome Treatment
1 86 YR Required Intervention