FDA Adverse Event Injury Summary report: N

ZENITH AAA ILIAC LEG GRAFT

MDR report key: 1033864 · Received April 25, 2008

Report

Report Number
1820334-2008-00230
Event Type
Injury
Date Received
April 25, 2008
Date of Event
March 28, 2008
Report Date
March 28, 2008
Manufacturer
COOK, INC.
Product Code
MIH
PMA / PMN Number
P020018
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVAL: COOK'S INSTRUCTIONS FOR USE DOES INDICATE THAT THE LONG-TERM PERFORMANCE OF THE ENDOVASCULAR GRAFTS HAS NOT YET BEEN ESTABLISHED. ALL PTS SHOULD BE ADVISED THAT ENDOVASCULAR TREATMENT REQUIRES LIFE-LONG, REGULAR F/U TO ASSESS THEIR HEALTH AND THE PERFORMANCE OF THEIR ENDOVASCULAR GRAFT. PTS WITH SPECIFIC CLINICAL FINDINGS (E.G., ENDOLEAKS, ENLARGING ANEURYSMS OR CHANGES IN THE STRUCTURE OR POSITION OF THE ENDOVASCULAR GRAFT) SHOULD RECEIVE ENHANCED F/U. THE DEVICE REMAINS IMPLANTED AND NO FILMS WERE PROVIDED TO ASSIST IN THIS INVESTIGATION. AN INTERNAL CLINICAL REVIEW INDICATED THAT A TYPE I AND TYPE II ENDOLEAK WERE PRESENT, DUE TO PT ANATOMY AND ANATOMICAL CHANGES OVER TIME.

Description of Event or Problem · 1

A MALE UNDERWENT INITIAL AAA REPAIR IN 2006. ONE MAIN BODY GRAFT AND TWO ILIAC LEG GRAFTS WERE PLACED. THE RIGHT ILIAC LIMB WAS RIGHT AT THE HYPOGASTRIC, AND IT HAD SEALED IN AN ILIAC ANEURYSM. HOWEVER, OVER TIME THE ILIAC ANEURYSM GREW AND THE AAA SAC BECAME PULSATILE. THE PHYSICIAN COULD NOT SEE ANY CONTRAST FILLING THE SAC, BUT COULD SEE A LITTLE AROUND THE LIMB. IN 2008, THE PHYSICIAN EXTENDED THE LIMB BY PLACING ANOTHER ILIAC GRAFT AS WELL AS COIL EMBOLIZED A LUMBAR AND THE HYPOGASTRIC. AFTER EXTENDING THE ILIAC LEG, THIS SEALED AND THE AAA SAC WAS NO LONGER PULSATILE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ZENITH AAA ILIAC LEG GRAFT MIH ENDOVASCULAR GRAFT MIH COOK, INC. NA 1687723

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention