FDA Adverse Event Injury Summary report: N

M2A-MAGNUM PF CUP 58ODX52ID

MDR report key: 10338268 · Received July 29, 2020

Report

Report Number
0001825034-2020-02976
Event Type
Injury
Date Received
July 29, 2020
Date of Event
January 15, 2019
Report Date
September 24, 2020
Manufacturer
ZIMMER BIOMET, INC.
Product Code
LPH
PMA / PMN Number
K101336
Removal / Correction Number
NI
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

REPORTED EVENT WAS CONFIRMED VIA MEDICAL RECORDS AND RADIOGRAPHS THAT WERE REVIEWED BY A HEALTH CARE PROFESSIONAL. REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED DEVIATIONS OR ANOMALIES DURING MANUFACTURING, THE DEVIATIONS OR ANOMALIES WOULD NOT HAVE ATTRIBUTED TO THE EVENT FOR THE INSERT. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: 157452, M2A-MAGNUM MOD HD SZ 52MM, 220570. 103209, TAPERLOC POR FMRL 17.5X155, 823640. 139268, M2A-MAGNUM 52-60MM TPR INS STD, 532050. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034 - 2020 - 02977, 0001825034 - 2020 - 02978, 0001825034 - 2020 - 02979. CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION, PRODUCT REMAINS IMPLANTED. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT UNDERWENT A LEFT HIP REVISION APPROXIMATELY 11 YEARS POST IMPLANTATION DUE TO ALTR, SWELLING, IN-VIVO CORROSION AND IMPLANT WEAR. DURING THE PROCEDURE THE BIOMET MAGNUM SLEEVE WAS FOUND TO BE COLD FUSED TO THE TRUNNION OF THE BIOMET TAPERLOC STEM. THICKENED PSEUDOCAPSULE WAS NOTED AND LARGE EFFUSION WHICH WAS NON-PURULENT. SOME BLACK CORROSIVE MATERIAL WAS NOTED WITHIN THE SYNOVIUM CONSISTENT THE METAL-ON-METAL BEARING AND METAL WEAR DEBRIS. THE BIOMET MAGNUM FEMORAL HEAD AND SLEEVE WERE NOTED TO BE COLD FUSED TO THE TRUNNION OF THE TAPERLOC STEM. MINIMAL CORROSION WAS NOTED AFTER APPROPRIATE CLEANING OF THE TRUNNION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
804167 M2A-MAGNUM PF CUP 58ODX52ID PROSTHESIS, HIP LPH ZIMMER BIOMET, INC. N/A 829560

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R