JP CHANNEL DRAIN 10 FR HUBLESS WITH TROCAR
Report
- Report Number
- 8030107-2020-00020
- Event Type
- Injury
- Date Received
- July 29, 2020
- Date of Event
- June 4, 2020
- Report Date
- July 29, 2020
- Manufacturer
- DEGANIA SILICONE LTD
- Product Code
- OTK
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
WE DID NOT RECEIVE THE ACTUAL DRAIN FOR THIS COMPLAINT FOR OUR EVALUATION. THE REASON FOR THE DRAIN TO ENTIRELY SLIP INTO THE WOUND MIGHT BE ITS BREAKAGE NEAR THE SURGICAL EXIT SITE OR THE DRAIN COULD GET INTO THE PATIENT BECAUSE IT WAS CUT DURING USE LEAVING SHORT PIECE OF 10 CM OUTSIDE THE SURGICAL EXIT SITE. WE EVALUATED THE BATCH HISTORY RECORDS OF THIS BATCH, PARTICULARLY THE RESULTS OF IN-PROCESS TENSILE STRENGS TESTS (TEAR TESTS) AND THE RESULTS WERE WELL ABOVE THE REQUIRED MINIMUM, SHOWING THAT THE DRAIN WAS NOT WEAKENED AND MOST LIKELY THE REASON FOR THE DRAIN TO GET INSIDE THE PATIENT, IS NOT A BREAKAGE/TEAR OF THE DRAIN FOLLOWING ANY MANUFACTURING ISSUE. WE CONCLUDE THAT MOST LIKELY THAT THE REASON FOR DRAIN TO SLIP INTO SURGICAL WOUND IS RELATED TO ITS USE (SHORT CUT) AND NOT TO MANUFACTURING PROBLEM.
PATIENT WENT FOR A PROCEDURE VIDEO LAPAROSCOPIC CHOLECYSTECTOMY ON THE (B)(6) 2020. CHANNEL DRAIN WAS INSERTED TO DRAIN THE WOUND. ON THE (B)(6) 2020, SURGEON MADE A CLINICAL DECISION TO CUT THE DRAIN AND LEFT A LENGTH OF 10 CM FROM THE SURGICAL EXIT SITE TO THE DRAINAGE BOTTLE. AT 10 PM ON THE SAME DAY, THE WARD NURSE NOTICED THAT THE DRAIN IS NO LONGER SEEN. AN X-RAY WAS ORDERED AND FOUND THE DRAIN IS INSIDE THE PATIENT'S BODY. A SURGERY WAS PERFORMED THE SAME DAY OF THE X-RAY TO REMOVE THE DRAIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 802236 | JP CHANNEL DRAIN 10 FR HUBLESS WITH TROCAR | JP CHANNEL DRAIN 10FR | OTK | DEGANIA SILICONE LTD | JP-2227 | P1836120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |