FDA Adverse Event Injury Summary report: Y

EDWARDS SAPIEN 3 ULTRA

MDR report key: 10335526 · Received July 28, 2020

Report

Report Number
2015691-2020-12758
Event Type
Injury
Date Received
July 28, 2020
Date of Event
May 29, 2020
Report Date
July 23, 2020
Manufacturer
EDWARDS LIFESCIENCES LLC
Product Code
NPT
PMA / PMN Number
P140031
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT TO ADD NOE STATEMENT IN B5 SECTION, AND PROVIDE D5 AND H6 INFORMATION.

Description of Event or Problem · 0

THIS REPORT SUMMARIZE <NOE> 2 </NOE> EVENTS OF AORTIC VALVE RE-INTERVENTION SERIOUS INJURY EVENTS FOR THE SAPIEN 3 ULTRA FOR (B)(6) OF 2020.

Additional Manufacturer Narrative · 1

EXEMPTION NUMBER E2016006, THIS REPORT SUMMARIZES 2 EVENTS OF AORTIC VALVE RE-INTERVENTION FOR THE SAPIEN 3 ULTRA VALVE IN THE AORTIC POSITION. THE ¿TIME TO EVENT¿ (TTE, IN DAYS) FOR THIS EVENT WAS 2.5. THE DEVICE IDENTIFICATION (DI) NUMBERS FOR EDWARDS SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE ARE: (B)(4). INTRA-PROCEDURAL AORTIC VALVE RE-INTERVENTION WILL TYPICALLY RESULT FROM VALVE MALPOSITION OR REGURGITATION (PVL OR CENTRAL, INCLUDING LEAFLET RESTRICTION). THESE CONDITIONS ARE KNOWN POTENTIAL RISKS ASSOCIATED WITH THE USE OF THE THV, DELIVERY SYSTEM, AND/OR ACCESSORIES. THERE ARE MULTIPLE PATIENT AND PROCEDURAL FACTORS THAT ALONE OR IN COMBINATION CAN CAUSE OR CONTRIBUTE TO AORTIC MALPOSITION, INCLUDING IMPROPER POSITIONING PRIOR TO DEPLOYMENT, POOR IMAGE INTENSIFIER ANGLE, POOR COAXIAL ALIGNMENT OF THE VALVE AND DELIVERY SYSTEM, SEVERE SEPTAL HYPERTROPHY, MINIMALLY OR BULKY/SEVERELY CALCIFIED AORTIC LEAFLETS, PRESERVED EJECTION FRACTION, SIGNIFICANT MITRAL ANNULAR CALCIFICATION (MAC), LOSS OF PACING CAPTURE, RAPID DEPLOYMENT, RELEASE OF STORED TENSION DURING DEPLOYMENT, AND MOVEMENT OF THE DELIVERY SYSTEM BY THE OPERATOR. DEPLOYMENT OF THE SAPIEN VALVE TOO AORTIC HAS THE POTENTIAL TO CONTRIBUTE TO SUBOPTIMAL COAPTATION OF THE SAPIEN VALVE LEAFLETS AND CAUSE CENTRAL AORTIC INSUFFICIENCY; IT CAN OBSTRUCT THE CORONARY OSTIA; AND LEAD TO EMBOLIZATION OF THE PROSTHESIS INTO THE ASCENDING AORTA. THE THV TRAINING MANUALS INSTRUCT THE OPERATOR ON PROPER POSITIONING AND DEPLOYMENT OF THE VALVE, INCLUDING ALL PROCEDURAL AND ANATOMICAL CONSIDERATIONS. PHYSICIANS ARE EXTENSIVELY TRAINED BY EDWARDS BEFORE THEY ARE QUALIFIED TO USE THE THV. TRAINING INCLUDES PATIENT SCREENING, DEVICE PREPARATION, APPROACH, DEPLOYMENT, IMAGING, PROCEDURE-SPECIFIC TRAINING MANUALS AND PROCTORED PROCEDURES. THE CORRECT ALIGNMENT AND POSITIONING OF THE DEVICE AT THE POINT OF DEPLOYMENT IS EMPHASIZED AS A KEY FACTOR TO THE PLACEMENT AND FIXATION OF THE DEVICE. OPERATORS ARE ALSO INSTRUCTED TO USE FLUOROSCOPY AS THE PRIMARY METHOD OF VISUALIZATION FOR POSITIONING AND DEPLOYMENT. IN PATIENTS WITH HIGH-RISK ANATOMICAL FEATURES FOR AORTIC MALPOSITION (I.E. MINIMAL LEAFLET CALCIFICATION, SEVERE SEPTAL HYPERTROPHY), BAV MAY PROVIDE INDICATION OF POTENTIAL BALLOON MOVEMENT DURING VALVE DEPLOYMENT. IN THIS CASE, SPECIFIC PROCEDURAL DETAILS ARE NOT AVAILABLE TO DETERMINE POTENTIAL CONTRIBUTING FACTORS. THE IFU AND TRAINING MANUALS HAVE BEEN REVIEWED AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS

Description of Event or Problem · 1

THV/TVT REGISTRY SUMMARY REPORTING FOR ADVERSE EVENT SUBMISSION FOR (B)(6) 2020 DATA EXTRACT FOR AORTIC SERIOUS INJURIES FOR THE SAPIEN 3 ULTRA VALVE. THIS REPORT SUMMARIZES 2 EVENTS OF AORTIC VALVE RE-INTERVENTION IN THE AORTIC POSITION SERIOUS INJURY EVENTS FOR THE SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE FOR (B)(6) 2020. THE AGE RANGE FOR THESE EVENTS IS FROM 81-86. THE BREAKDOWN FOR GENDER IS AS FOLLOWS: 1 FEMALE AND 1 MALE. (B)(6) 2020 DATA EXTRACT INCLUDES DATA PROVIDED BY ACC FOR Q4 2019 (OCTOBER 1 ¿ DECEMBER 31).