FDA Adverse Event Injury Summary report: N

MIRA LUX MULTIFLEX TURBINE 635B

MDR report key: 1033371 · Received April 24, 2008

Report

Report Number
8010493-2008-00007
Event Type
Injury
Date Received
April 24, 2008
Date of Event
February 26, 2008
Report Date
March 27, 2008
Manufacturer
KALTENBACH & VOIGT GMBH
Product Code
EKX
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

PT WAS SENT TO EMERGENCY AND BUR WAS REMOVED WITH NO PROBLEMS. NO MEDICATION WAS GIVEN AND PT WAS RELEASED THE SAME DAY. EVAL OF DEVICE FOUND THAT THE BEARINGS WERE WORN AND GRITTY, THERE WAS DAMAGE TO THE BACK END CAP, AND MEDIUM DEBRIS WAS PRESENT. MAINTENANCE ISSUES WERE DISCUSSED. HANDPIECE WAS REPLACED.

Description of Event or Problem · 1

BUR CAME OUT OF HANDPIECE AND LODGED IN PT'S THROAT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MIRA LUX MULTIFLEX TURBINE 635B DENTAL HANDPIECE AND ACCESSORIES EKX KALTENBACH & VOIGT GMBH 635B

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R