FDA Adverse Event
Injury
Summary report: N
MIRA LUX MULTIFLEX TURBINE 635B
MDR report key: 1033371
·
Received April 24, 2008
Report
- Report Number
- 8010493-2008-00007
- Event Type
- Injury
- Date Received
- April 24, 2008
- Date of Event
- February 26, 2008
- Report Date
- March 27, 2008
- Manufacturer
- KALTENBACH & VOIGT GMBH
- Product Code
- EKX
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
PT WAS SENT TO EMERGENCY AND BUR WAS REMOVED WITH NO PROBLEMS. NO MEDICATION WAS GIVEN AND PT WAS RELEASED THE SAME DAY. EVAL OF DEVICE FOUND THAT THE BEARINGS WERE WORN AND GRITTY, THERE WAS DAMAGE TO THE BACK END CAP, AND MEDIUM DEBRIS WAS PRESENT. MAINTENANCE ISSUES WERE DISCUSSED. HANDPIECE WAS REPLACED.
Description of Event or Problem · 1
BUR CAME OUT OF HANDPIECE AND LODGED IN PT'S THROAT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MIRA LUX MULTIFLEX TURBINE 635B | DENTAL HANDPIECE AND ACCESSORIES | EKX | KALTENBACH & VOIGT GMBH | 635B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |