FDA Adverse Event Injury Summary report: N

ARTHREX ECLIPSE HUMERAL HEAD, 49/18

MDR report key: 10333402 · Received July 28, 2020

Report

Report Number
1220246-2020-01981
Event Type
Injury
Date Received
July 28, 2020
Date of Event
July 14, 2020
Report Date
August 13, 2020
Manufacturer
ARTHREX, INC.
Product Code
QHQ
UDI-DI
00888867060067
PMA / PMN Number
K183194
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

COMPLAINT NOT CONFIRMED, NO ABNORMALITY WAS FOUND ON THE DEVICE THAT MAY HAVE CONTRIBUTED TO THE EVENT. THE CAUSE IS UNDETERMINED.

Additional Manufacturer Narrative · 1

THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD NOT BE DETERMINED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE INFORMATION AVAILABLE AND WITHOUT DEVICE EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A TOTAL SHOULDER ARTHROPLASTY USING THE ECLIPSE SYSTEM ON (B)(6) 2020. UPON REVIEW OF THE POST OPERATIVE X-RAY, ON POST-OP DAY 1, IT WAS NOTICED THAT THE SHOULDER WAS DISLOCATED. A SECOND X-RAY WAS PERFORMED THE SAME DAY AND IT WAS CONFIRMED THAT THE SHOULDER WAS IN FACT DISLOCATED. THE PATIENT WAS NOT ABLE TO REMEMBER WHEN OR HOW THE SHOULDER MAY HAVE DISLOCATED. A REVISION TSA-HUMERAL HEAD EXCHANGE SURGERY WAS SCHEDULED AND PERFORMED ON (B)(6) 2020 (POST-OP DAY 1). IT WAS NOTED DURING THE REVISION THAT THE SUBSCAPULARIS REPAIR WAS STILL IN TACT BUT HAD TO BE TAKEN DOWN TO GET THE SHOULDER RELOCATED. AFTER THE SHOULDER WAS RELOCATED THE ECLIPSE HUMERAL HEAD, AR-9349-18 (LOT 18.0147) WAS REMOVED AND THE ENTIRE SHOULDER JOINT WAS EVALUATED FOR ANY POSSIBLE PROBLEM. THE SURGEON THEN PROCEEDED TO TRIAL THREE DIFFERENT HUMERAL HEAD SIZES AND SETTLED ON THE MOST STABLE HUMERAL HEAD WHICH WAS AN AR-9349-20 (LOT 1735002). A SPEEDBRIDGE TECHNIQUE WAS PERFORMED TO REPAIR THE SUBSCAPULARIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
795145 ARTHREX ECLIPSE HUMERAL HEAD, 49/18 TOTAL SHOULDER ARTHROPLASTY SYSTEM QHQ ARTHREX, INC. ARTHREX ECLIPSE HUMERAL HEAD, 49/18 18.01427 00888867060067

Patients

Seq Age Sex Outcome Treatment
1 Other