SICAT IMPLANT V2.0
Report
- Report Number
- 3006098230-2020-00001
- Event Type
- Malfunction
- Date Received
- July 28, 2020
- Date of Event
- July 1, 2020
- Report Date
- October 30, 2020
- Manufacturer
- SICAT GMBH & CO. KG
- Product Code
- LLZ
- PMA / PMN Number
- K192348
- Removal / Correction Number
- 3006098230-2020-001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
Narratives
SICAT IMPLANT V2.0 IS A STANDALONE SOFTWARE APPLICATION. THE DEVICE PROBLEM WAS ANALYZED AND IT WAS FOUND THAT IMPLANT PLANNING DATA EXPORTED FROM SICAT IMPLANT ARE NOT CORRECT FOR ONE SPECIFIC EXPORT FORMAT. A CODING ERROR WAS DETECTED CAUSING THE DEVICE PROBLEM. IMMEDIATE ACTION HAS BEEN TAKEN WITHIN THE LICENSING SYSTEM OF SICAT IMPLANTV 2.0 TO PREVENT NEW CUSTOMERS/USERS FROM ACTIVATING THE APPLICATION. A FIELD SAFETY NOTICE HAS BE SEND OUT TO THE USERS AND CONSIGNEES OF SICAT IMPLANT V2.0. THE DEVICE WILL BE CORRECTED IN THE FIELD BY DISTRIBUTING AND APPLYING A SOFTWARE PATCH TO REMOVE THE PROBLEM. EVALUATION RESULTS AND EVALUATION REPORT WILL BE PROVIDED IN A FOLLOW-UP REPORT.
SICAT IMPLANT IS A STANDALONE SOFTWARE APPLICATION, I.E. A RADIOLOGICAL VISUALIZATION SOFTWARE FOR DIAGNOSIS AND DENTAL IMPLANT PLANNING. A DENTIST FOUND THAT IMPLANT PLANNING DATA EXPORTED FROM SICAT IMPLANT ARE NOT CORRECT FOR ONE SPECIFIC EXPORT FORMAT. IF IT WERE TO RECUR, THE ERROR IN THE EXPORT FORMAT MIGHT CAUSE DENTISTS TO MANUFACTURE (OR HAVE MANUFACTURED) SURGICAL GUIDES WHICH ARE INCORRECT. USING THOSE GUIDES MIGHT CAUSE DRILLING INTO THE JAW TO BE EITHER NOT DEEP ENOUGH OR TO DEEP. A DENTIST MIGHT HAVE TO STOP A SURGICAL PROCEDURE. OR THE DRILLING WHEN TO DEEP MIGHT CAUSE DAMAGE TO STRUCTURES WITHIN THE JAW WHICH NEED INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 799309 | SICAT IMPLANT V2.0 | RADIOLOGICAL VISUALIZATION SOFTWARE FOR DIAGNOSIS AND DENTAL IMPLANT PLANNING | LLZ | SICAT GMBH & CO. KG | SICAT IMPLANT V2.0 | V2.0 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR | DENTSPLY SIRONA CERCEC SYSTEM (INLAB18 SOFTWARE) |