PRECISION LINK
Report
- Report Number
- 2954323-2008-01551
- Event Type
- Malfunction
- Date Received
- April 22, 2008
- Date of Event
- March 24, 2008
- Report Date
- April 22, 2008
- Manufacturer
- ABBOTT DIABETES CARE INC.
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- UNKNOWN
Narratives
THERE IS A KNOWN MALFUNCTION WITH THE PRECISION LINK SOFTWARE THAT CAN LEAD TO INCORRECT TRENDING OF RESULTS. THIS OCCURS WHEN RESULTS OBTAINED ON A METER WITH INCORRECT DATE AND TIME, ARE UPLOADED TO A COMPUTER WITH PRECISION LINK SOFTWARE. CUSTOMERS AND RETAILERS HAVE BEEN INFORMED.
A CUSTOMER REPORTED RECEIVING AN ERROR 6 ON THEIR PRECISION XTRA BLOOD GLUCOSE METER. IT WAS THEN ADDITIONALLY IDENTIFIED BY ADC CUSTOMER SVC THAT THE DATE AND TIME SETTINGS IN THEIR METER WERE NOT PROPERLY SET, AND THEY REPORTED TO BE A USER OF THE PRECISION LINK DATA MGMT SYS. THE CUSTOMER REPORTED "FEELING SLOW AND RUN DOWN" AND SELF-TREATING WITH 2 TUBES OF GLUCOSE AND ORANGE JUICE TO COUNTERACT THE SYMPTOMS. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRECISION LINK | DATA MANAGEMENT SYSTEM | NBW | ABBOTT DIABETES CARE INC. | NA | 42498 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |