FDA Adverse Event Malfunction Summary report: N

LIFEPAK 9P DEFIBRILLATOR/MONITOR/PACEMAKER

MDR report key: 1033054 · Received April 23, 2008

Report

Report Number
3015876-2008-00369
Event Type
Malfunction
Date Received
April 23, 2008
Date of Event
January 25, 2008
Report Date
January 25, 2008
Manufacturer
PHYSIO-CONTROL, INC
Product Code
LDD
PMA / PMN Number
K902288
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE POWER SUPPLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD TURN ON/OFF BY ITSELF. THERE WAS NO PT USE ASSOCIATED WITH THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 9P DEFIBRILLATOR/MONITOR/PACEMAKER LDD PHYSIO-CONTROL, INC 9P NA

Patients

Seq Age Sex Outcome Treatment
1 NA