DIMENSION CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2008-00040
- Event Type
- Malfunction
- Date Received
- April 22, 2008
- Date of Event
- March 26, 2008
- Report Date
- March 26, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS, INC.
- Product Code
- MMI
- PMA / PMN Number
- K944093
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
NO FURTHER EVAL OF THE DEVICE IS REQUIRED. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS CARRY IN FROM THE R1 PROBE. THE MALFUNCTION WAS RESOLVED AFTER THE CUSTOMER PERFORMED MAINTENANCE WITH GUIDANCE FROM A SIEMENS HEALTHCARE DIAGNOSTICS INC TECHNICAL SOLUTION CENTER REP. THE INSTRUMENT IS PERFORMING WITHIN SPECS.
A FALSELY ELEVATED TROPONIN I RESULT OF 1.27 NG/ML WAS OBTAINED ON A PT SAMPLE. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS RETESTED AND A RESULT OF 0.04 NG/ML WAS OBTAINED. PT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE OF THE FALSELY ELEVATED TROPONIN I RESULT WAS CARRYING FROM THE R1 PROBE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | MMI | SIEMENS HEALTHCARE DIAGNOSTICS, INC. | RXLRH | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |