FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1032839 · Received April 22, 2008

Report

Report Number
2531779-2008-00223
Event Type
Injury
Date Received
April 22, 2008
Date of Event
March 22, 2008
Report Date
March 22, 2008
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE PATIENT'S FATHER REPORTED, THAT THE PATIENT INADVERTENTLY ADMINISTERED EXCESS INSULIN BY PRIMING THE PUMP, WHILE ATTACHED TO THE INFUSION SET. THE PUMP USER GUIDE INSTRUCTS USERS TO DISCONNECT FROM THE INFUSION SET PRIOR TO PERFORMING A PRIME. ADDITIONALLY, THE PUMP NOTIFIES THE USER TO DISCONNECT FROM THE INFUSION SET THREE TIMES DURING THE REWIND/PRIME SEQUENCE. THE PATIENT HAS CONTINUED TO USE THE PUMP WITH NO REPORTED SUBSEQUENT EVENTS. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.

Description of Event or Problem · 1

THE PATIENT RECEIVED EMERGENCY ROOM TREATMENT FOR HYPOGLYCEMIA. THE PATIENT INADVERTENTLY ADMINISTERED EXCESS INSULIN BY PRIMING THE PUMP WHILE ATTACHED TO THE INFUSION SET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANIMAS 2020

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention