FDA Adverse Event Malfunction Summary report: N

ABBOTT COMMANDER PPC

MDR report key: 1032667 · Received April 22, 2008

Report

Report Number
1628664-2008-00101
Event Type
Malfunction
Date Received
April 22, 2008
Date of Event
March 25, 2008
Report Date
March 25, 2008
Manufacturer
ABBOTT MANUFACTURING INC
Product Code
JQW
PMA / PMN Number
K897158
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

OTHER: GAS SHOCK. THE ABBOTT COMMANDER PARALLEL PROCESSING CENTER (LN 06208-86) HAD A DAMAGED COVER SPRING ON THE LEFT SIDE OF THE PROCESSOR, SO THE LID WOULD NOT STAY OPEN. NO INJURIES WERE REPORTED AS A RESULT OF THE DAMAGED SPRING. A NEW SPRING WAS SENT TO THE CUSTOMER FOR INSTALLATION. THE CUSTOMER INSTALLED THE SPRING AND VISUALLY VERIFIED THE PROPER OPERATION. A REVIEW OF THE COMPLAINT HISTORY FOR ABBOTT COMMANDER PARALLEL PROCESSING CENTER (LN 06208-86) WAS PERFORMED FOR THE TIME PERIOD OF 2008. NO RELATED COMPLAINTS WERE IDENTIFIED. THIS IS A FINAL REPORT.

Description of Event or Problem · 1

THE ABBOTT FIELD SERVICE REP WAS AT THE ACCOUNT SITE AND DISCOVERED THE SPRING ON THE LEFT LID COVER OF THE COMMANDER PPC WAS DAMAGED AND NEEDED TO BE REPLACED. THE LID WAS NOT HOLDING UP AS EXPECTED BECAUSE OF THE DAMAGED SPRING. NO INJURY WAS REPORTED DUE TO COMMANDER PPC LID NOT STAYING OPEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ABBOTT COMMANDER PPC AUTOMATED SYSTEM FOR THE PROCESSING OF EIAS JQW ABBOTT MANUFACTURING INC NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA GAS SPRING-MOD. TOP COVER