DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2020-077691
- Event Type
- Malfunction
- Date Received
- July 27, 2020
- Date of Event
- July 24, 2020
- Report Date
- February 3, 2021
- Manufacturer
- DEXCOM, INC.
- Product Code
- QBJ
- PMA / PMN Number
- DEN170088
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). B5: DESCRIBE EVENT OR PROBLEM - CORRECTION. H2: CORRECTION.
(B)(4). B5: DESCRIBE EVENT OR PROBLEM - ADDITIONAL D10: DEVICE AVAILABLE FOR EVALUATION - ADDITIONAL H2: ADDITIONAL INFORMATION/DEVICE EVALUATION H3: DEVICE EVALUATED BY MANUFACTURER - ADDITIONAL H6: EVENT PROBLEM AND EVALUATION CODES - ADDITIONAL.
AN INTERNAL VISUAL INSPECTION WAS PERFORMED FAILED DUE TO FOREIGN OBJECT DAMAGE ON THE SPEAKER DUST COVER.
THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE AND BOOT TESTS WERE PERFORMED AND PASSED. FUNCTIONAL TESTS WERE PERFORMED AND FAILED THE SPEAKER TESTS. A "TRY IT" SOUND TEST WAS PERFORMED AND FAILED. AN INTERNAL VISUAL INSPECTION WAS PERFORMED FAILED DUE TO FOD ON SPEAKER DIAPHRAGM. PICTURES WERE TAKEN. THE RECEIVER LOG WAS DOWNLOADED AND REVIEWED FINDING NO ERROR RELATED TO THE COMPLAINT. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE FOREIGN OBJECT DAMAGE. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
(B)(4).
IT WAS REPORTED THAT LOW AUDIO OUTPUT OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.