FDA Adverse Event Malfunction Summary report: N

WALLABY 3 PHOTO THERAPY UNIT

MDR report key: 1032563 · Received March 27, 2008

Report

Report Number
1032563
Event Type
Malfunction
Date Received
March 27, 2008
Date of Event
March 10, 2008
Report Date
March 27, 2008
Manufacturer
RESPIRONICS, INC.
Product Code
LBI
Product Problem
Yes
Report Source
User Facility report
Reporter Location
PA, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

PARENTS REPORT THAT THEY WERE AWAKENED BY A SHATTERING NOISE, "A HUGE BANG" "LIKE A WINDOW SHATTERING." THE PARENTS WENT TO THE ROOM WHERE THEIR INFANTS WERE AND OBSERVED GLASS ON THE FLOOR (THE INFANTS WERE RECEIVING PHOTOTHERAPY). THE YELLOW LIGHT ON WALLABY UNIT WAS FLASHING AND THE GREEN LIGHT WAS ON. THE UNIT APPEARED TO BE OPERATING CORRECTLY AND THE SECOND BULB HAD TURNED ON AUTOMATICALLY AND PHOTOTHERAPY WAS BEING DELIVERED. THE PARENTS REPORTED UNPLUGGING THE UNIT AND CLEANING UP THE GLASS. NO REPORT OF INJURY. THE UNIT WAS PLUGGED BACK IN TO CONTINUE THERAPY. THE PARENTS CALLED THE HOSPITAL TO PROVIDE NOTIFICATION OF THE EVENT. THE HOME WAS VISITED BY A STAFF MEMBER. THE INVOLVED WALLABY UNIT WAS RECOVERED AND RETURNED TO THE MANUFACTURER FOR EVALUATION.MANUFACTURER RESPONSE FOR PHOTOTHERAPY, SYSTEM, INFANT, WALLABY 3 PHOTO THERAPY UNIT: PER A REPORT FORWARDED FROM THE COMPANY - UNABLE TO CONFIRM COMPLAINT OF "BULB EXPLODED." THERE IS NO EVIDENCE, SHARDS, OR PARTS INTERNAL OF UNIT. HOWEVER THERE IS A BULB ENCLOSURE INSTALLED WITH A MISSING ELEMENT. UNIT ALSO APPEARS TO HAVE BEEN DROPPED. THE FRONT PANEL ENCLOSURE IS CRACKED/MARRED/DAMAGED. THE OUTER METAL ENCLOSURE IS DENTED ON THE LEFT REAR CORNER. REPLACED BULBS WITH KNOWN GOOD BULBS AND RAN FOR 24 HOURS TOTAL (SWITCHING BOTH). NO OPERATIONAL ISSUES WERE FOUND/NOTED. QA WILL HANG ONTO AND ARCHIVE THE SUSPECT BULB FOR A PERIOD OF ONE YEAR. UNIT IS SERVICEABLE AND WILL BE RETURNED IN SERVICE FOR REPAIR AFTER INVESTIGATION COMPLETED AND UNIT RELEASED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WALLABY 3 PHOTO THERAPY UNIT PHOTOTHERAPY SYSTEM, INFANT LBI RESPIRONICS, INC. * *

Patients

Seq Age Sex Outcome Treatment
1 5 DA