FDA Adverse Event Malfunction Summary report: N

FLEXLAB

MDR report key: 10325080 · Received July 27, 2020

Report

Report Number
3010825766-2020-00006
Event Type
Malfunction
Date Received
July 27, 2020
Date of Event
June 24, 2020
Report Date
February 25, 2021
Manufacturer
INPECO SA
Product Code
CEM
PMA / PMN Number
K121012
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE INITIAL REPORT (3010825766-2020-00006) HAS BEEN SUBMITTED ON JULY 27TH, 2020. ADDITIONAL INFORMATION (FEBRUARY 22ND, 2021): PER INPECO'S REQUEST, THE DAMAGED COMPONENT WAS SENT BACK TO FURTHER INVESTIGATE THE CAUSE OF THE MALFUNCTION. THE CAPACITORS OF THE POWER SUPPLY APPEAR TO BE SWOLLEN. THE DAMAGE OF THE CAPACITORS CAUSED THE LEAK OF THE ELECTROLYTE AND THE SHORT CIRCUIT. THE FAILURE CAN BE A CONSEQUENCE OF THE AGEING AND OF THE NORMAL WEAR AND TEAR SINCE THE COMPONENT WAS IN USE FOR 7 YEARS. THE DESIGN OF INPECO DEVICE HAS BEEN EVALUATED APPROPRIATE SINCE THE AUTOMATION SYSTEM IS CERTIFIED FOR ELECTRICAL SAFETY AND THE INVOLVED POWER SUPPLY IS INSTALLED INSIDE A FIRE ENCLOSURE WHICH AVOIDS THE FLAMES PROPAGATION IN CASE OF MALFUNCTION. THE INVESTIGATION DID NOT HIGHLIGHT THE NEED OF ANY DESIGN CHANGE. AFTER THE DAMAGED POWER SUPPLY REPLACEMENT, THE AUTOMATION SYSTEM IS CONTINUING TO WORK ACCORDING TO SPECIFICATIONS. NO FURTHER ACTIONS ARE FORESEEN. THE CODES IN SECTION H6 HAVE BEEN UPDATED IN ACCORDANCE TO THE INVESTIGATION CONCLUSIONS.

Additional Manufacturer Narrative · 1

AFTER THE DAMAGED POWER SUPPLY REPLACEMENT, THE AUTOMATION SYSTEM IS PERFORMING ACCORDING TO SPECIFICATIONS. THE INVESTIGATION IS STILL ONGOING TO DETERMINE THE ROOT CAUSE.

Description of Event or Problem · 1

THE DISTRIBUTOR NOTIFIED INPECO ABOUT THE FAILURE OF A 24VDC, 20A POWER SUPPLY INSTALLED IN THE FLEXLAB AUTOMATION SYSTEM TRACK. THE OPERATOR WAS NOTIFIED OF THE ISSUE BY A POWER SUPPLY ERROR (1202). THE OPERATOR DID NOT REPORT ANY SMOKE OR FLAMES. WHEN THE COMPONENT WAS REPLACED, THE FSE NOTICED VISIBLE SIGNS OF BURNING ON AN INTERNAL CAPACITOR. THE POWER SUPPLY WAS CONTAINED WITHIN THE TRACK BEHIND A PLEXIGLAS COVER AND THE METAL TRACK COVER, SO THE OPERATOR COULD NOT COME IN DIRECT CONTACT WITH THE DAMAGED COMPONENT. THE DISTRIBUTOR REPORTED THAT THE FAILURE DID NOT CAUSE ANY INJURY OR IMPACT TO PATIENT RESULT PROCESSING. THE EVENT DID NOT CAUSE THE DAMAGE OF ANY OTHER TRACK COMPONENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
793123 FLEXLAB LABORATORY AUTOMATION SYSTEM CEM INPECO SA

Patients

Seq Age Sex Outcome Treatment
1