FDA Adverse Event Other Summary report: N

GSPM TABLE

MDR report key: 1032482 · Received February 12, 2008

Report

Report Number
1932056-2007-00006
Event Type
Other
Date Received
February 12, 2008
Date of Event
December 13, 2007
Report Date
December 27, 2007
Manufacturer
MEDICAL POSITIONING, INC.
Product Code
FQO
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

INCIDENT RESULTED FROM WELD FAILURE. WHILE LESS THAN 0.02% WELDS HAVE FAILED, THE INCIDENCE OF FAILURE IN 2004 IS HIGHER THAN IN OTHER YEARS. FOR THAT REASON, ALL CUSTOMERS WITH MODELS WHO HAVE REPORTED WELD FAILURE, AND ARE FROM BATCHES OF STEEL USED IN 2004, WILL BE CONDUCTED AND AN UPGRADE WILL BE INSTALLED TO ENSURE NO FAILURES IN THE FUTURE.

Description of Event or Problem · 1

PAIN MANAGEMENT TABLE "BROKE" WITH PATIENT ON TABLE. TABLE WENT FROM HORIZONTAL TO 45 DEGREES HEAD DOWN. PATIENT SLID OFF TABLE. THE PATIENT DESCRIBED THE EVENT AS CAUSING A "JOLT" TO HIS RIGHT ARM AND SHOULDER. HE WAS TAKEN TO THE EMERGENCY ROOM FOR EVAL, BUT THE EXTENT OF HIS INJURIES IS UNKNOWN. IT WAS DISCOVERED THAT THE LOWER TRENDELENBURG ACTUATOR MOUNT HAD BROKEN AWAY FROM THE FRAME. THIS CAUSED THE HEAD END OF THE TABLE TO FALL SUDDENLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GSPM TABLE GENERAL SURGERY & PAIN MANAGEMENT TABLE FQO MEDICAL POSITIONING, INC. 60326

Patients

Seq Age Sex Outcome Treatment
1 49 YR Other