FDA Adverse Event Other Summary report: N

CONTIPLEX CONTINUOUS PLEXUSTRAY W/ ACCUFUSER

MDR report key: 1032469 · Received April 16, 2008

Report

Report Number
2523676-2008-00021
Event Type
Other
Date Received
April 16, 2008
Date of Event
March 11, 2008
Report Date
April 11, 2008
Manufacturer
B. BRAUN MEDICAL, INC.
Product Code
CAZ
PMA / PMN Number
K030830
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE IN THE INCIDENT WAS NOT RETURNED TO THE MANUFACTURER TO BE EVALUATED. HOWEVER, TWO PHOTOGRAPHS OF THE KNOTTED CATHETER WERE RETURNED, ALONG WITH AN UNUSED REPRESENTATIVE CATHETER SAMPLE. THERE WERE NO VISUAL MANUFACTURING DEFECTS NOTED ON THE RETURNED UNUSED CATHETER. THE PHOTO DEPICTED THE CATHETER LOOPED AND KNOTTED IN SEVERAL PLACES. INCIDENTS OF THIS NATURE CAN GENERALLY BE ATTRIBUTED TO EXCESSIVE CATHETER THREADING WHICH INCREASES THE LIKELIHOOD OF ENTANGLEMENT. BASED ON THE PHOTOGRAPH ANALYSIS AND THE ADDITIONAL INFORMATION PROVIDED BY THE FACILITY, IT APPEARS THAT THE CATHETER MAY HAVE BEEN THREADED TOO FAR INTO THE AXILLARY AREA. HOWEVER, WITHOUT THE ACTUAL SAMPLE, NO SPECIFIC CONCLUSIONS CAN BE DRAWN. THE UNUSED SAMPLE AND THE PHOTOGRAPHS ALONG WITH ALL AVAILABLE INFORMATION HAS BEEN FORWARDED TO THE MANUFACTURER B. BRAUN FOR FURTHER EVALUATION.

Description of Event or Problem · 1

AS REPORTED BY THE USER FACILITY: FEMALE PATIENT. PROCEDURE PERFORMED. EXPERIENCED NO DIFFICULTY PLACING CATHETER. REQUIRED SURGICAL REMOVAL BY OPENING UP THE AXILLARY AREA. CATHETER WAS FOUND TO BE KNOTTED. ADDITIONAL INFORMATION PROVIDED BY THE FACILITY INDICATED THE SURGERY TO REMOVE THE CATHETER WAS PERFORMED WITHOUT INCIDENT AND THE PATIENT HAS NOT SUFFERED ANY ADDITIONAL ADVERSE SEQUELA ASSOCIATED WITH THIS INCIDENT. IT WAS ALSO REPORTED THE PHYSICIAN ADVANCED THE CATHETER 6CM, DURING THE INITIAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTIPLEX CONTINUOUS PLEXUSTRAY W/ ACCUFUSER CONTIPLEX CONTINUOUS NERVE BLOCK KIT CAZ B. BRAUN MEDICAL, INC. NA 60858609

Patients

Seq Age Sex Outcome Treatment
1 Other