FDA Adverse Event
Malfunction
Summary report: N
HUTIV TABLE ASSY STD W/COLOR
MDR report key: 1032305
·
Received April 22, 2008
Report
- Report Number
- 1518293-2008-00160
- Event Type
- Malfunction
- Date Received
- April 22, 2008
- Date of Event
- March 28, 2008
- Report Date
- March 28, 2008
- Manufacturer
- LIEBEL-FLARSHEIM CO.
- Product Code
- MMZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
LIEBEL FLARSHEIM FIELD SERVICE ENGINEER REPORT. FSE REPLACE TOWER SIDE TABLE SUPPORT RAIL AND TOWER SIDE THUMB SCREW LATCH ON VOIDING STOOL DUE TO DAMAGED THREADS ON BOTH ITEMS. UNIT FUNCTIONAL AND RETURNED TO SERVICE.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT A FEMALE WAS UNDERGOING A UROLOGY PROCEDURE FOR INCONTINENCE. SHE REPORTED THAT THEY ATTACHED THE VOIDING STOOL TO THE UROLOGY TABLE. WITH THE PATIENT SITTING ON THE STOOL THEY STARTED TO RAISE THE PATIENT INTO POSITION WHEN THE VOIDING STOOL AND PATIENT TILTED SIDEWAY. THE CUSTOMER REMOVED THE PATIENT FROM THE STOOL. SHE REPORTED THAT THERE WERE NO INJURIES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HUTIV TABLE ASSY STD W/COLOR | UROLOGY TABLE | MMZ | LIEBEL-FLARSHEIM CO. | HUT TABLE | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |