FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK 20 DEFIBRILLATOR/MONITOR
MDR report key: 1032291
·
Received April 21, 2008
Report
- Report Number
- 3015876-2008-00366
- Event Type
- Malfunction
- Date Received
- April 21, 2008
- Date of Event
- January 14, 2008
- Report Date
- January 14, 2008
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- LDD
- PMA / PMN Number
- K012274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE POWER SUPPLY ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EVALUATED THE REPLACED ASSEMBLY AND DETERMINED THAT ROOT CAUSE FOR THE REPORTED FAILURE WAS THE ELECTRICALLY SHORTED FET TRANSISTORS, DESIGNATORS Q1, Q2 AND Q6.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD NOT POWER ON WITH AC POWER. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 20 DEFIBRILLATOR/MONITOR | LDD | PHYSIO-CONTROL, INC. | 20 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |