ACS RX ESPRIT CORONARY DILATATION CATHETER (PERFUSION)

MDR report key: 1032290 · Received April 21, 2008

Report

Report Number
2024168-2008-00328
Event Type
Malfunction
Date Received
April 21, 2008
Date of Event
March 21, 2008
Report Date
March 24, 2008
Manufacturer
ABBOTT VASCULAR-CARDIAC THERAPIES
Product Code
LOX
PMA / PMN Number
P810046
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

RESULTS AND CONCLUSION SUMMATION - PRODUCT PERFORMANCE ENGINEERING REVIEWED THE INCIDENT INFORMATION. FACTORS THAT CAN CONTRIBUTE TO BALLOON RUPTURE INCLUDE, BUT ARE NOT LIMITED TO, BALLOON DAMAGE DURING MANUFACTURING, MATERIALS, INTERACTIONS WITH OTHER DEVICES, PATIENT ANATOMY, LESION CALCIFICATION, LESION TORTUOSITY, OR INSUFFICIENT PREPARATION PRIOR TO USE. THE LESION WAS MODERATELY CALCIFIED, WHICH COULD HAVE CONTRIBUTED TO THE BALLOON RUPTURE. ALSO, THE ESPRIT WAS USED TO EXPAND A STENT, WHICH HAD NOT FULLY DEPLOYED, WHICH SUGGESTS THAT THE STENT STRUTS COULD HAVE CONTRIBUTED TO THE BALLOON RUPTURE. HOWEVER, WITHOUT A RETURNED DEVICE FOR ANALYSIS, A DEFINITE ROOT CAUSE FOR THE BALLOON RUPTURE CANNOT BE DETERMINED.

Description of Event or Problem · 1

REPORTING STATUS: MALFUNCTION. REPORTING RATIONALE: BALLOON RUPTURE IS LIKELY TO CAUSE OR CONTRIBUTE TO PATIENT INJURY. DEVICE ISSUE: BALLOON RUPTURE. IT WAS REPORTED THAT THE LESION WAS PRE-DILATED AND ANOTHER COMPANY'S STENT WAS PRESSURIZED TO DEPLOY. WHEN THE PRESSURE REACHED 10 ATM, A PERFORATION OCCURRED. THE STENT DELIVERY SYSTEM WAS REMOVED AND THE ESPRIT WAS ENGAGED FOR HEMOSTASIS. DURING THE FIRST INFLATION, THE BALLOON OF THE ESPRIT RUPTURED AT 6 ATM. IT WAS REPLACED BY ANOTHER COMPANY'S BALLOON. THE PATIENT WAS TRANSFERRED TO SURGERY TO TREAT THE PERFORATION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACS RX ESPRIT CORONARY DILATATION CATHETER (PERFUSION) LOX ABBOTT VASCULAR-CARDIAC THERAPIES NA 6042051

Patients

Seq Age Sex Outcome Treatment
1 UNK STENT CYPHER 2.5X18| DIL CATH: VENT 2.0X15| PRESTIGE 3.5X20