FDA Adverse Event Malfunction Summary report: N

VOYAGER RX CORONARY DILATATION CATHETER

MDR report key: 1032289 · Received April 21, 2008

Report

Report Number
2024168-2008-00330
Event Type
Malfunction
Date Received
April 21, 2008
Date of Event
March 25, 2008
Report Date
March 26, 2008
Manufacturer
ABBOTT VASCULAR-CARDIAC THERAPIES
Product Code
LOX
PMA / PMN Number
P810046
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

RESULTS AND CONCLUSION SUMMATION - PRODUCT PERFORMANCE ENGINEERING REVIEWED THE INCIDENT INFORMATION. FACTORS THAT CAN CONTRIBUTE TO BALLOON RUPTURE INCLUDE, BUT ARE NOT LIMITED TO, BALLOON DAMAGE DURING MANUFACTURING, MATERIALS, INTERACTIONS WITH OTHER DEVICES, PATIENT ANATOMY, LESION CALCIFICATION, LESION TORTUOSITY, OR INSUFFICIENT PREPARATION PRIOR TO USE. REPORTEDLY, THE LESION TREATED IN THE PROCEDURE WAS HEAVILY CALCIFIED, WHICH COULD HAVE CONTRIBUTED TO DAMAGING THE SURFACE OF THE BALLOON SUCH THAT UPON INFLATION IS RUPTURED. HOWEVER, WITHOUT A RETURNED DEVICE FOR ANALYSIS, A DEFINITE ROOT CAUSE FOR THE BALLOON RUPTURE CANNOT BE DETERMINED.

Description of Event or Problem · 1

REPORTING STATUS: MALFUNCTION. REPORTING RATIONALE: BALLOON RUPTURE IS LIKELY TO CAUSE OR CONTRIBUTE TO PATIENT INJURY. DEVICE ISSUE: BALLOON RUPTURE. IT WAS REPORTED THAT DURING THE FIRST INFLATION WITH THE RX VOYAGER, THE BALLOON RUPTURED AT 6 ATM. ANOTHER COMPANY'S BALLOON CATHETER WAS USED TO COMPLETE THE PROCEDURE. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VOYAGER RX CORONARY DILATATION CATHETER LOX ABBOTT VASCULAR-CARDIAC THERAPIES NA 7040233

Patients

Seq Age Sex Outcome Treatment
1 UNK