FDA Adverse Event
Malfunction
Summary report: N
ORTHO PROVUE
MDR report key: 1032012
·
Received April 21, 2008
Report
- Report Number
- 1056600-2008-00120
- Event Type
- Malfunction
- Date Received
- April 21, 2008
- Date of Event
- March 25, 2008
- Report Date
- April 21, 2008
- Manufacturer
- MICRO TYPING SYSTEMS, INC.
- Product Code
- KSZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
A POSSIBLE ROOT CAUSE WAS DETERMINED. AN OCD FIELD ENGINEER ARRIVED ON SITE AND REPLACED THE BENT PROBE TO RETURN THE INSTRUMENT TO EXPECTED OPERATION. THIS CUSTOMER HAS NOT LOGGED ANY COMPLAINTS AGAINST THIS ANALYZER SINCE THE REPAIR.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE PROBE ON THE ORTHO PROVUE ANALYZER WAS BENT AND DRIPPING FLUID. THE CUSTOMER INDICATED THAT THE REAGENTS WERE CONTAMINATED. TESTING WAS ABORTED, AND NO ERRONEOUS RESULTS WERE REPORTED. PROBE DRIP MAY LEAD TO DILUTION OF SAMPLE / REAGENT, CARRY OVER AND / OR CROSS CONTAMINATION AND ERRONEOUS RESULTS WHICH COULD LEAD TO TRANSFUSION OF INCOMPATIBLE BLOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ORTHO PROVUE | AUTOMATED BLOOD GROUPING & ANTIBODY TEST SYSTEM | KSZ | MICRO TYPING SYSTEMS, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |