FDA Adverse Event
Malfunction
Summary report: N
FREESTYLE FREEDOM
MDR report key: 1031978
·
Received April 18, 2008
Report
- Report Number
- 2954323-2008-01523
- Event Type
- Malfunction
- Date Received
- April 18, 2008
- Date of Event
- March 22, 2008
- Report Date
- April 18, 2008
- Manufacturer
- ABBOTT DIABETES CARE INC., USA
- Product Code
- LFR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER'S PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE FILED ONCE INVESTIGATION RESULTS ARE AVAILABLE.
Description of Event or Problem · 1
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR FREESTYLE FREEDOM BLOOD GLUCOSE MONITOR. CUSTOMER REPORTED RECEIVING READINGS OF 52 MG/DL, 221 MG/DL, 226 MG/DL AND SECOND 221 MG/DL WITHIN 10 MINUTES. ALL TESTS WERE PERFORMED ON THE FINGER. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONES SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FREEDOM | BLOOD GLUCOSE MONITORING SYSTEM | LFR | ABBOTT DIABETES CARE INC., USA | NA | 0729405 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |