TUBE, GASTROINTESTINAL (AND ACCESS)
Report
- Report Number
- 2523003-2008-00040
- Event Type
- Malfunction
- Date Received
- April 17, 2008
- Date of Event
- February 10, 2008
- Report Date
- March 25, 2008
- Manufacturer
- C.R. BARD, INC (BASD)
- Product Code
- KNT
- PMA / PMN Number
- K915837
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE COMPLAINT OF A BREAK IN THE RETENTION DOME IS CONFIRMED, USER RELATED. UPON RECEIPT, A PARTIAL SEMI-CIRCULAR BREAK IN THE RETENTION DOME WAS OBSERVED. THE DULL AND ROUNDED VENEER IDENTIFIED AT THE DOME SITE ARE TRAITS THAT ARE INDICATIVE OF EXCESSIVE FORCE THAT WAS APPLIED DURING THE REMOVAL OF THIS DEVICE. THE LACK OF ANY ASSOCIATED DAMAGE SUGGESTS THE TEAR IN THE DOME WAS CAUSED BY STRETCHING THE DOME MATERIAL BEYOND ITS ELASTIC LIMITS DURING REMOVAL. THE COMPLAINANT INDICATES THE COMPLAINT SAMPLE HAD BEEN IN PLACE FOR APPROX 138 DAYS PRIOR TO THE COMPLAINT INCIDENT. THE BROKEN RETENTION DOME WAS NOT RETURNED FOR EVAL. GROSS VISUAL AND MICROSCOPIC EXAMINATIONS SHOW NO EVIDENCE OF A DEFECT OR DEFORMITY RELATED TO THE MANUFACTURING PROCESS. A CHR OF LOT # HURJ0256 SHOWED NO OTHER PROD COMPLAINTS FROM THIS LOT NUMBER.
A BREAK IN A RETENTION DOME DURING TRACTION REMOVAL. THE INDWELLING PERIOD WAS 138 DAYS. THE DOME PORTION WAS REMOVED WITH BOWEL MOVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TUBE, GASTROINTESTINAL (AND ACCESS) | KNT | C.R. BARD, INC (BASD) | HURJ0256 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |