FDA Adverse Event Malfunction Summary report: N

CONTOUR TEST STRIPS (50)

MDR report key: 1031896 · Received April 15, 2008

Report

Report Number
1826988-2008-00392
Event Type
Malfunction
Date Received
April 15, 2008
Date of Event
March 21, 2008
Report Date
March 21, 2008
Manufacturer
BAYER HEALTHCARE LLC
Product Code
NBW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER CALLED FOR HELP WITH HER METER. WHILE TROUBLESHOOTING, SHE PERFORMED CONTROL TESTS AND RECEIVED A RESULT OF 212 MG/DL. THE NORMAL CONTROL RANGE WAS 103-142 MG/DL. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVAL. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS (50) BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE LLC 7080D 7FC3D02

Patients

Seq Age Sex Outcome Treatment
1 UNK