FDA Adverse Event Other Summary report: N

ALARIS SYSTEM PUMP MODULE

MDR report key: 1031757 · Received April 18, 2008

Report

Report Number
2016493-2008-00018
Event Type
Other
Date Received
April 18, 2008
Date of Event
March 19, 2008
Manufacturer
CARDINAL HEALTH ALARIS PRODUCTS
Product Code
FRN
PMA / PMN Number
K950419
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B) (4). (B) (4). THIS REPORT WAS FILED BY THE MFR.

Description of Event or Problem · 1

INFUSION WAS STARTED AROUND 2 AM ON 3/19; AT 4 AM RN FOUND BAG EMPTY. INFUSION WAS HYDROMORPHONE, 1 MG/ML IN A 100ML BAG. THE ENTIRE BAG INFUSED IN 2 HOURS, NOT THE 12 HOURS IT WAS INTENDED TO RUN AS IT WAS SUPPOSED TO BE PROGRAMMED AT 8ML/HR. PT WAS OPIATE TOLERANT ON LARGE DOSES OF MULTIPLE NARCOTICS, AND THERE WAS NO EFFECT ON THE PATIENT; NO NARCOTIC REVERSAL AGENT WAS ADMINISTERED. BAG WAS IN A "CLAMSHELL" TYPE OF SECURITY DEVICE THAT WOULD NOT NECESSARILY PREVENT DIVERSION BUT IS DESIGNED TO MAKE IT DIFFICULT TO GET TO THE VOLUME; CUSTOMER RULED OUT POSSIBILITY OF DIVERSION. EVENT LOGS FROM THE PCU WERE REVIEWED BY CARDINAL. THERE WERE THREE LVPS ATTACHED TO THE PCU AT THE TIME OF THE EVENT. ONE WAS INFUSING AT A RATE OF 100 ML/HR, ONE WAS INFUSING A PRIMARY AT A RATE OF 40 ML/HR. TWO HOURS BEFORE THE TIME IN QUESTION A RATE OF 8 ML/HR AND A VTBI OF 100 ML WERE PROGRAMMED INTO THE THIRD PUMP, AND 2 HOURS LATER THIS PUMP WAS SELECTED, THE SECONDARY INFUSION OPTION WAS ACTIVATED, THE SECONDARY INFUSION RATE WAS SELECTED AS 100 ML /HR AND THE VTBI AT 102 ML. IT IS BELIEVED THAT A SECONDARY INFUSION PROGRAM MEANT FOR THE SECOND PUMP WAS PROGRAMMED INTO THE THIRD PUMP; BASED ON THIS LOG REVIEW THE CAUSE FOR THE REPORTED OVER INFUSION IS BELIEVED TO BE USER PROGRAMMING ERROR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALARIS SYSTEM PUMP MODULE FRN, INFUSION PUMP FRN CARDINAL HEALTH ALARIS PRODUCTS 8100 NA

Patients

Seq Age Sex Outcome Treatment
1 Other ALARIS SYSTEM PC UNIT (B) (4)| ALARIS SYSTEM PUMP MODULES (B) (4) & (B) (4)