LIFEPAK 20 DEFIBRILLATOR/MONITOR

MDR report key: 1031701 · Received April 18, 2008

Report

Report Number
3015876-2008-00360
Event Type
Malfunction
Date Received
April 18, 2008
Date of Event
March 20, 2008
Report Date
March 20, 2008
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
LDD
PMA / PMN Number
K012274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND OBSERVED THAT IT WOULD NOT OPERATE ON AC POWER. PHYSIO REPLACED THE POWER SUPPLY ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

ACCORDING TO THE REPORTER, THE DEVICE WILL NOT OPERATE ON AC AND IT WILL NOT CHARGE THE BATTERY. THERE WERE NO ADVERSE AFFECTS REPORTED AS A RESULT OF THE DEVICE USE.