Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE POWER SUPPLY ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. FURTHER EVALUATION OF THE REPLACED ASSEMBLY DETERMINED THAT ROOT CAUSE FOR THE FAILURE TO OPERATE ON BATTERY POWER WAS AN ELECTRICALLY LEAKY FILTER, FL4.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD NOT OPERATE ON BATTERY POWER. THERE WAS NO PT USE ASSOCIATED WITH THE REPORTED EVENT.