FDA Adverse Event Malfunction Summary report: N

LAB BOVIE PENCIL

MDR report key: 1031489 · Received April 17, 2008

Report

Report Number
MW5006290
Event Type
Malfunction
Date Received
April 17, 2008
Date of Event
March 26, 2008
Report Date
April 17, 2008
Manufacturer
COVIDIEN VALLEY
Product Code
GEI
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
SC, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

DISPOSABLE BOVIE PENCIL CAUSED FIRE HAZARD. DISPOSABLE BOVIE LOT# 131269 IGNITED AND CHARRED RAYTEC SPONGE. ALL OTHER EQUIPMENT IN THE OR WAS CHECKED AND THE DISPOSABLE BOVIE WAS IDENTIFIED AS THE CAUSE. THIS ITEM WAS CHANGED TWICE, A NEW DISPOSABLE BOVIE PENCIL WITH A DIFFERENT LOT NUMBER WAS USED WITHOUT PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAB BOVIE PENCIL GEI COVIDIEN VALLEY E2515H 131269

Patients

Seq Age Sex Outcome Treatment
1 71 YR