FDA Adverse Event Malfunction Summary report: N

LIGACLIP ENDOSCOPIC MULTIPLE CLIP APPLIER

MDR report key: 103148 · Received July 2, 1997

Report

Report Number
1527736-1997-01475
Event Type
Malfunction
Date Received
July 2, 1997
Report Date
July 1, 1997
Manufacturer
ETHICON ENDO-SURGERY, INC.
Product Code
GDO
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE WHICH WAS RETURNED TO CO WITH PRODUCT INQUIRY #973441. VISUAL INSPECTIONS & RESULTS: CLIP IN JAW, NO; DAMAGED JAWS, YES; DAMAGED CUTTER, N/A; DAMAGED FEED BAR, NO; DAMAGED FLOOR, NO; DAMAGED HANDLE SHROUDS, NO; DAMAGED OTHER, NO; DAMAGED TIP SHROUDS, NO; DAMAGED TRIGGER, NO; DAMAGED TUBE, NO AND DAMAGED WELD SEAMS, NO. FUNCTIONAL TESTS & RESULTS: ANTIBACKUP FUNCTIONAL, YES; FIRING: FEED CONFORM, YES; FIRING: FORM CONFORM, NO; JAWS: HOLD CLIP, YES; JAWS: INSIDE WIDTH AT TIPS, .222 AND LOCKOUT FUNCTIONAL, YES. ANALYSIS CONCLUSION: BASED UPON THE INQUIRY INFO RECEIVED AND THE VISUAL FUNCTIONAL RESULTS, IT WAS CONCLUDED THAT THE JAWS HAD BECOME DAMAGED AND WOULD NOT FORM THE CLIPS PROPERLY. NO CONCLUSION COULD BE REACHED AS TO HOW THE DAMAGE TO THE JAWS OCCURRED. IF THE JAWS ARE CLOSED OVER A HARD OBJECT, THE JAWS CAN BECOME DAMAGED AND WILL NOT FORM THE CLIPS PROPERLY. EACH INSTRUEMNT IS EVALUATED DURING THE ASSEMBLY PROCESS TO ENSURE THE CLIPS FEED AND FORM PROPERLY. CO STRIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMPROVE CO'S PRODUCTS.

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE WAS USED DURING AN UNKNOWN PROCEDURE. IT WAS REPORTED BY THE AFFILIATE THAT THE CLIPS WERE MALFORMED (GAPPED). THERE WAS NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGACLIP ENDOSCOPIC MULTIPLE CLIP APPLIER ENDOSCOPIC CLIP APPLIER GDO ETHICON ENDO-SURGERY, INC. NA K46C88

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other