GUNTHER TULIP JUGULAR VENA CAVA FILTER SET
Report
- Report Number
- 1820334-2020-01350
- Event Type
- Injury
- Date Received
- July 23, 2020
- Report Date
- July 23, 2020
- Manufacturer
- COOK INC
- Product Code
- DTK
- UDI-DI
- 10827002330174
- PMA / PMN Number
- K072240
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE PREVIOUS MEDWATCH REPORT WAS SUBMITTED BY WILLIAM COOK EUROPE UNDER MANUFACTURER REPORT REFERENCE# 3002808486-2020-00596. (B)(6). ADDITIONAL INFORMATION PROVIDED DETERMINED THAT THIS DEVICE WAS MANUFACTURED BY COOK INC. WITH THE SUBMISSION OF THIS INITIAL MEDWATCH REPORT, COOK INC. INFORMS THAT ALL FUTURE SUBMISSIONS REGARDING THIS COMPLAINT WILL BE HANDLED UNDER THE MANUFACTURER REPORT REFERENCE # REFERENCED IN THIS INITIAL MEDWATCH REPORT. BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN OR UNAVAILABLE. INVESTIGATION: INVESTIGATION IS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. THE REPORTED ALLEGATIONS HAVE BEEN FURTHER INVESTIGATED BASED ON THE INFORMATION PROVIDED TO DATE. THE FOLLOWING ALLEGATIONS HAVE BEEN INVESTIGATED: VENA CAVA (VC) PERFORATION, NUMBNESS, LEG PAIN, LIMITED MOBILITY, DEPRESSION, ANXIETY. FILTER INTERACTS WITH IVC WALL, E.G. PENETRATION/PERFORATION/EMBEDMENT. THIS MAY BE EITHER SYMPTOMATIC OR ASYMPTOMATIC. POTENTIAL CAUSES MAY INCLUDE IMPROPER DEPLOYMENT; AND (OR) EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IN-SITU FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER). POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: TRAUMA TO ADJACENT STRUCTURES, VASCULAR TRAUMA, VENA CAVA PERFORATION, VENA CAVA PENETRATION. UNKNOWN IF THE REPORTED NUMBNESS, LEG PAIN, LIMITED MOBILITY, DEPRESSION, AND ANXIETY ARE DIRECTLY RELATED TO THE FILTER AND UNABLE TO IDENTIFY A CORRESPONDING FAILURE MODE AT THIS POINT IN TIME. 20 DEVICES IN LOT. NO RELEVANT NOTES ON WORK ORDER. THE PRODUCT IS MANUFACTURED AND INSPECTED ACCORDING TO SPECIFICATIONS. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE.
PATIENT ALLEGEDLY RECEIVED AN IMPLANT VIA THE RIGHT INTERNAL JUGULAR VEIN DUE TO DEEP VEIN THROMBOSIS (DVT) AND PULMONARY EMBOLISM (PE). PATIENT IS ALLEGING VENA CAVA PERFORATION. THE PATIENT FURTHER ALLEGES "NUMBNESS, LEG PAIN, UNABLE TO WALK ANY AMOUNT OF DISTANCE," AS WELL AS DEPRESSION AND ANXIETY. REPORT FROM CT (COMPUTED TOMOGRAPHY): "IVC FILTER IN PLACE. THE PRONGS ABUT THE IVC DISTAL TO THE RENAL VEINS. THE IVC PRONGS APPEAR TO EXTEND BEYOND THE EXPECTED WALL OF THE IVC"
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 781968 | GUNTHER TULIP JUGULAR VENA CAVA FILTER SET | DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR | DTK | COOK INC | G33017 | 4050766 | 10827002330174 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Life Threatening |