LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM
Report
- Report Number
- 2024601-2008-00197
- Event Type
- Malfunction
- Date Received
- April 18, 2008
- Date of Event
- January 22, 2008
- Report Date
- March 20, 2008
- Manufacturer
- COSTA RICA
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
TAPER II. VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS OF THE DEVICE NOTED DAMAGE FROM A SHARP INSTRUMENT, CONSISTENT WITH SURGICAL DAMAGE, TO THE BAND TUBING (THIS IS THE PORTION OF TUBING LOCATED BETWEEN THE STAINLESS STEEL CONNECTOR AND THE BAND, NOT THE STAINLESS STEEL CONNECTOR AND THE PORT). THERE WAS NO INDICATION OF WEAR RELATED DAMAGE TO THIS PORTION OF THE LAP-BAND SYSTEM RETURNED. PERFORMANCE TESTS INDICATE NO LEAKAGE AT THE ACCESS PORT OR ACCESS PORT TUBING. THERE IS NO OTHER INFO AVAILABLE AT THIS TIME FROM THE SURGEON TO DETERMINE THE CAUSE OF THE ALLEGED EVENT. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."
REPORTED AS THE DOCTOR WAS UNABLE TO REMOVE THE SAME AMOUNT OF FLUID PUT IN. (LEAK)
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM | DEVICE FOR TREATMENT OF MORBID OBESITY | LTI | COSTA RICA | NA | 1304054 |