FDA Adverse Event Malfunction Summary report: N

LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM

MDR report key: 1031427 · Received April 18, 2008

Report

Report Number
2024601-2008-00197
Event Type
Malfunction
Date Received
April 18, 2008
Date of Event
January 22, 2008
Report Date
March 20, 2008
Manufacturer
COSTA RICA
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TAPER II. VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS OF THE DEVICE NOTED DAMAGE FROM A SHARP INSTRUMENT, CONSISTENT WITH SURGICAL DAMAGE, TO THE BAND TUBING (THIS IS THE PORTION OF TUBING LOCATED BETWEEN THE STAINLESS STEEL CONNECTOR AND THE BAND, NOT THE STAINLESS STEEL CONNECTOR AND THE PORT). THERE WAS NO INDICATION OF WEAR RELATED DAMAGE TO THIS PORTION OF THE LAP-BAND SYSTEM RETURNED. PERFORMANCE TESTS INDICATE NO LEAKAGE AT THE ACCESS PORT OR ACCESS PORT TUBING. THERE IS NO OTHER INFO AVAILABLE AT THIS TIME FROM THE SURGEON TO DETERMINE THE CAUSE OF THE ALLEGED EVENT. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."

Description of Event or Problem · 1

REPORTED AS THE DOCTOR WAS UNABLE TO REMOVE THE SAME AMOUNT OF FLUID PUT IN. (LEAK)

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM DEVICE FOR TREATMENT OF MORBID OBESITY LTI COSTA RICA NA 1304054

Patients

Seq Age Sex Outcome Treatment
1