FULL FACE MASK
Report
- Report Number
- 9611451-2008-00197
- Event Type
- Malfunction
- Date Received
- April 18, 2008
- Date of Event
- March 3, 2008
- Report Date
- March 3, 2008
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- MNT
- PMA / PMN Number
- K060044
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE IS EN ROUTE TO THE MFR. THE INVESTIGATION WAS CONDUCTED BASED ON THE EVENT DESCRIPTION AND PREVIOUS INVESTIGATIONS ON SIMILAR COMPLAINTS. A LOT CHECK HAS REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR THIS LOT NUMBER. THE IMPROPER FITTING OF THE MASK MAY HAVE CONTRIBUTED TO THE SILICONE SEAL SEPARATING FROM THE BASE OF THE MASK. THE RT040 MASK IS RELATIVELY NEW TO THE MARKET AND MANY USERS HAVE BEEN CONVERTING FROM A PREVIOUSLY USED COMPETITOR'S DEVICE TO THE RT040 AND AS A RESULT MAY NOT BE PROFICIENT WITH THE SIZING AND FITTING OF THE RT040. CONCLUSIONS: FISHER & PAYKEL HEALTHCARE HAS IMPLEMENTED AN IN-SERVICE EDUCATION PROGRAM TO FURTHER ENSURE THAT HEALTHCARE PRACTITIONERS FULLY UNDERSTAND THE PROPER FITTING OF THE RT040 MASK. THIS PROGRAMME HAS BEEN AND CONTINUES TO BE ROLLED OUT TO CUSTOMERS IN THE UNITED STATES. IN ADDITION, WE HAVE INTRODUCED AN ADDITIONAL SILICONE ADHESIVE STEP (TO APPLY ADHESIVE BETWEEN THE SILICONE FACIAL SEAL AND THE PLASTIC MASK BASE) IN OUR MFG PROCESS WHICH IS AIMED AT REDUCING THE POTENTIAL FOR SEPARATION TO OCCUR. WE HAVE RECEIVED SIMILAR COMPLAINTS AND ARE CURRENTLY TRENDING THIS AT AN OCCURRENCE RATE OF APPROX 0.0531% WORLDWIDE FOR THE PAST YEAR.
A HEALTHCARE FACILITY REPORTED TO OUR DISTRIBUTOR THAT AN RT040L FULL FACE MASK FELL APART IN THE PACKAGE. NO PT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FULL FACE MASK | MNT | FISHER & PAYKEL HEALTHCARE, LTD. | RT040L | 071116 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |