FDA Adverse Event Malfunction Summary report: N

COSYCOT INFANT WARMER

MDR report key: 1031410 · Received April 18, 2008

Report

Report Number
9611451-2008-00141
Event Type
Malfunction
Date Received
April 18, 2008
Date of Event
January 11, 2008
Report Date
January 17, 2008
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
FMT
PMA / PMN Number
K971695
Removal / Correction Number
Z-0485-2007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

REPLACE THE CRACKED LEGS ON THE INFANT WARMER ELEVATOR BASE. AN INVESTIGATION WAS CARRIED OUT BASED ON THE EVENT DESCRIPTION AND RESULTS OF PREVIOUS INVESTIGATIONS ON SIMILAR COMPLAINTS, AS THE COMPLAINT DEVICE HAD NOT BEEN RETURNED. RESULTS: CRACKING LEGS OF THE COSYCAT INFANT WARMER, DUE TO EXCESSIVE WEIGHT LOADING, IS A KNOWN PROBLEM. TOTAL WEIGHT OF THE DEVICE, INCLUDING ACCESSORIES AND PT, EXCEEDING 115 KG MAY, UNDER CERTAIN CIRCUMSTANCES, CAUSE CRACKS IN THE PLASTIC LEGS' BASE AND DAMAGE TO THE BASE ELECTRIC ELEVATOR MECHANISM. CONCLUSION: WE HAVE AN OPEN CAPA TO ADDRESS THIS ISSUE. THE CORRECTIVE ACTION, INFORMING THE USERS OF THE MAXIMUM WEIGHT FOR THE COSYCOT INFANT WARMER AND TO INSPECT THE BASES OF THEIR COSYCOTS, HAS BEEN COMPLETED IN THE USA, BUT IS STILL ON GOING IN OTHER COUNTRIES. THIS CORRECTIVE ACTION HAS BEEN NOTIFIED TO THE DISTRICT OFFICE.

Description of Event or Problem · 1

A HOSPITAL IN ANOTHER COUNTRY, REPORTED THAT A CRACK WAS FOUND ON THE LEG OF THE IW932AEK COSYCAT INFANT WARMER ELEVATOR BASE. NO PT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COSYCOT INFANT WARMER FMT FISHER & PAYKEL HEALTHCARE, LTD. IW932AEK

Patients

Seq Age Sex Outcome Treatment
1